background image

 

 

 
Lid release 

Once the run has been completed and the rotor come to a stop, the lid will open 
automatically. If the lid does not open automatically, press the lid button. Note 
that the lid button will not operate while the rotor is spinning. 

WARNING: Do not attempt to open the lid of any centrifuge until the rotor 
has come to a complete stop. 

 
In the event of a power failure or malfunction, it may be necessary to open the lid 
manually. 
1. Disconnect the power cord from the wall socket. 
2. Remove the plastic plug, located on the left side of the unit, below the quick 
button. 
3. Pull the cord (attached to the plug) to open the lid lock manually. 

 
Lid lock 

The centrifuge can be started only with the lid securely closed. When the rotor 
begins to accelerate, the lid lock indicator light turns on and the lid button 
becomes inoperable. Do not attempt to open the lid until the lid lock indicator 
turns off. At the end of the run, the lid will automatically open. 

 
Speed selection 

The speed (rpm) can be selected to 14,000rpm with the knob labeled “speed”. 
The scale is directly proportional to the speed - a setting of 9 corresponds to 
9,000rpm, a setting of 13 corresponds to a speed of 13,000rpm, etc. 
 

Selection of operating time and momentary operation 

Operation of the centrifuge begins when the timer knob is turned clockwise to set 
a run time. For run times less than 5 minutes, turn the knob clockwise past the 
halfway point and then counterclockwise to the desired time. For run times longer 
than 5 minutes, turn the knob clockwise to the desired time.  
 
When the preselected time expires, the centrifuge will stop automatically. To stop 
the centrifuge prior to the expiration of set time, turn the timer knob to the zero 
position.  
 
The centrifuge may be operated manually by pressing and holding the quick 
button. The centrifuge will continue to run as long as the button is depressed.  
 
Some models are equipped with a timer that includes a hold position. Continuous 
operation of these models may be achieved by turning the timer knob firmly to 
the left. The centrifuge will continue to operate until the knob is turned to the zero 
position. 
 
Note: The timer knob may be turned in either direction during operation of the 
centrifuge without damage to the timer mechanism. 

 

 

Summary of Contents for Spectrafuge 16M

Page 1: ...Lit Code M00001 Rev 2 August 2016 Spectrafuge 16M Microcentrifuge User Manual C0160 C0160 100V C0160 230V...

Page 2: ...illage may harm unit NEVER put hands in the rotor area unless the rotor is completely stopped NEVER move the centrifuge while the rotor is spinning NEVER use solvents or flammables near this or other...

Page 3: ...tors and accessories Rotor maintenance Mounting and securing the angle rotors Removing the rotor Overloading rotors Operation 5 Closing the lid Lid release Lid lock Speed selection Selection of operat...

Page 4: ...mpaired Do not operate the centrifuge if any of the following conditions exist The centrifuge has not been installed properly The centrifuge is partially dismantled Service has been attempted by unaut...

Page 5: ...table laboratory bench cabinet etc To guarantee sufficient ventilation ensure that the centrifuge has at least 15cm 6 inches of free space on all sides including the rear The centrifuge should not be...

Page 6: ...ting and securing the angle rotor Remove the rotor screw from the motor shaft by turning the screw counterclockwise Clean the motor shaft and the rotor mounting hole see figures 1 2 Place the rotor on...

Page 7: ...4 Figure 1 Chamber and motor shaft Figure 2 Bottom of angle rotor Figure 3 Loading the rotor...

Page 8: ...igher density it is necessary to reduce the speed Failure to reduce the speed may result in damage to the rotor and centrifuge The revised maximum speed can be calculated with the following formula Re...

Page 9: ...labeled speed The scale is directly proportional to the speed a setting of 9 corresponds to 9 000rpm a setting of 13 corresponds to a speed of 13 000rpm etc Selection of operating time and momentary...

Page 10: ...The rotor should be cleaned after each use When spinning 8 samples containing phenol or phenol chloroform the rotor should be cleaned immediately after use Disinfection Should a spill of infectious ma...

Page 11: ...lose lid correctly Possible reason No speed or time has been selected Solution Set speed and or time Where to call Should you have any questions about the Spectrafuge 16M or its accessories please cal...

Page 12: ...l Force chart can be used to determine g values To use this chart find the radius value on the radius scale Place the edge of a ruler on the value Place the right side edge of the ruler on the speed s...

Page 13: ...s symbol identifies a Protective Earth PE terminal which is provided for connection of the supply system s protective earth green or green yellow conductor EQUIPMENT DISPOSAL EUROPEAN REGULATIONS Acco...

Page 14: ...paint or finish Claims for transit damage should be filed with the transportation carrier ALL WARRANTIES INCLUDING THE IMPLIED WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE LIMI...

Page 15: ...NOTES...

Page 16: ...9299560000 31 Mayfield Ave Edison NJ 08837 P 732 417 0700 F 732 417 1750 Toll Free 888 522 6381 labnetinfo corning com www labnetinternational com...

Reviews: