CHAPTER 3 OPERATION GUIDE 

 
 
There  are  eight  menu  selections  for  the  MultiGene  Mini  Thermal  Cycler.  In  this 
chapter, we will explain operation, for example, how to edit, access, modify, delete, 
or run a protocol file and how to set parameters.  
 

Warning!

  If an abnormal sound or display appears on start up or if a failure error 

during the Self Test mode, Turn off the power immediately and contact 
Labnet or your distributor  

 
 

Note: 

If the number of samples is less than the number of wells in the block, 
the sample tubes should be evenly distributed over the block, in order 
to ensure that the heated lid will evenly press on the tops of the sample 
tubes. This also ensures that both the block’s load and the temperature 
distribution to each tube are even.  

 
 

Note: 

Before closing the cover, turn the knob counter-clockwise according to 
the  knob  diagram  until  there  is  a  "click....."  Please  note  that  the  knob 
can  be  tightened  by  turning  it  clockwise,  After  closing  the  cover  but 
before  power  on,  turn  the  knob  clockwise  until  you  hear  a  click,  then 
turn the knob one more turn. 
When the knob is turned clockwise or counter-clockwise, you will hear a 
"click" sound, To open pull the release latch with your finger; and turn 
the knob counter-clockwise. 

 

1

  How to Run a Program 

 
Switch  on  the  power,  the  MultiGene  Mini  Thermal  Cycler  will  beep  4  times.  The 
screen  will  display  “

and  the  instrument  will  carry  on  self  test. 

The self test takes about 1~2 minutes. If the self test is successful, the screen will 
display the main Menu.  
 
On  the  left  top  corner  of  the  main 
screen the Hot lid status is displayed.  
 
The  center  of  the  screen  displays 
Labnet, the software version number, 
block model, default file, default user, 
control mode and sample volume. 
 

  Press

  “File”

  to  enter  file  list 

interface. 

  Press 

“System”

 to enter system 

parameter setting menu. 

Main Screen

 

Hotlid=020C                        2009-04-01 10:05:00 
 

Labnet 

           

MultiGene Mini 

         

Version:       2001-1.0 

          Default User:    Labnet  
          Default File:     1 
          Module Select:   24*0.2 
          Control Mode:   Block  

           

 

 
 

    File                     System                     Run  

                      

Summary of Contents for MultiGene Mini

Page 1: ...MultiGene Mini Thermal Cycler TC020 24 TC020 24 230V TC050 18 TC050 18 230V Operation Manual Ver 1 2 Labnet International...

Page 2: ...3 1 Description 3 2 Keyboard Diagram 4 3 Key Functions 4 4 Inspection before Power up 5 5 Power on Procedures 5 CHAPTER 3 OPERATION GUIDE 6 1 How to Run a Program 6 1 1 How to edit a program 7 1 2 How...

Page 3: ...0Hz 220 240V 200W 50 60Hz Note Before powering on please confirm that the above operating conditions are satisfied The unit must be plugged into a grounded outlet 2 Transportation and Storage Conditio...

Page 4: ...of cycles 99 Max Segments within a cycle 9 Max temperature time 99 min 59 sec Max incubation and in situ time 99 hour 59 min Program storage capacity 100 5 Software Functions File editing and saving...

Page 5: ...er on 1 Description Heated lid Block LCD Display Keyboard RS232 interface Power switch Power connecter Fuse Socket Note 1 The fuse specifications are as follows 250V 2 5A 125V 5A 5 20mm The fuse shoul...

Page 6: ...g file parameter setting Up Down moving the cursor one position up or down during file parameter setting Enter press it to accept the present parameter settings displayed on the screen Function press...

Page 7: ...e power socket 5 Power on Procedures Turn on the power i e turn the power switch to position 1 the MultiGene Mini Thermal Cycler will beep 4 times and the LCD will display This means the device is sel...

Page 8: ...an be tightened by turning it clockwise After closing the cover but before power on turn the knob clockwise until you hear a click then turn the knob one more turn When the knob is turned clockwise or...

Page 9: ...s below If the current Control Mode is Block mode press Accept button to enter file running interface If the current Control Mode is Tube mode press Accept button a dialog box will be displayed to inp...

Page 10: ...tected press Run a dialog box will be displayed to input the password Press Back to go back the file list interface Press Accept to confirm the password After pressing Run in the file list interface a...

Page 11: ...by pressing Up Down and Right Left Press the keys to modify parameter settings Press CE to change Press Seg to add a segment In every segment you can set temperature Temp duration Time Ramping rate R...

Page 12: ...normally saved File has been saved Please continue another operation will appear Press Enter to close the reminder window If the file name was already in use the system will display Rename or continu...

Page 13: ...ess Stop and Do you want to stop running will be displayed Select F1 or F3 You can then select either Stop or Start Press Pause and the display will show Pause state now Press F3 to continue running Y...

Page 14: ...trol Mode stands for default control model after power on including two optional modes named Block and Tube Block mode offers basic temperature control during the course of cooling heating only Compar...

Page 15: ...temperature which allows the sample to reach the set temperature faster and will help the stay at the set temperature for a longer period Please refer to the following figure relation between block te...

Page 16: ...r bad temperature control accuracy Cooling module failure Contact Labnet or your distributor Cooling module failure 3 Significant change in cooling speed or unavailable temperature below ambient Fan f...

Page 17: ...oken there is a short circuit or the fan is not working 4 Error Message when running Please contact the distributor and Temperature sensor No 1 is disconnected Block sensor No 1 reports an open circui...

Page 18: ...ATER LID J2 4_HEATER CPU PCB4 BYQ5041041000000_CPU J13 1_HV J13 2_HV J13 3_VIN5 J13 4_HOTLID J13 5_FAN J13 6_PWM J13 7_HEAT J13 8_COOL J3_CPU_JTAG_IDC20 J10_KEY IN J5_LCD_24P J9_CPLD_JTAG_IDC10 J16 1_...

Page 19: ...P J9_CPLD_JTAG_IDC10 J16 1_SEN HOTLID J16 2_SEN HOTLID J15 1_SEN FAN J15 2_SEN FAN J15 3_SEN BLOCK J15 4_SEN BLOCK VB1_SUBD9_RS232C J17 1_H J17 2_H L U5 1 2 3 4 5 6 L I N P E N I N N L P E DCFAN U4 1...

Page 20: ...date 9 Time Time HH MM SS 00 00 00 23 59 59 hour minute second 10 Default File The default file name Max length is 11 characters 11 Default User The default user name Max length is 11 characters Block...

Page 21: ...19 Image 2 Image1 block reagent block reagent...

Page 22: ...ts in original material or workmanship This warranty does not cover motor brushes fuses light bulbs batteries or damage to paint or finish Claims for transit damage should be filed with the transporta...

Reviews: