Labnet MultiGene Mini Operation Manual Download Page 14

 

 

12 

1.2 How to Set System Parameters 

 
 
In the main screen press “

System

” to 

enter  the  system  parameter  settings 
menu.  Press  “

Enter

”  to  move  the 

cursor to the next item. This operation 
is  followed  by  the  symbol  “ ”.  Press 

Up/Down

” to scroll to an option, and 

information  can  be  entered  by  using 
the  keyboard.  Press  “

Right/Left” 

to 

move cursor’s position.

 

  Press  “

Back

”  to  go  back  to  the 

main menu. 

  Press “

Save

” to save parameters. 

 
 

Note:

 

When  setting  a  parameter,         

press the “Save” button, otherwise the changes will be lost. 
 

1.2.1 “Date” and “Time” indicate the system time and date. 

1.2.2 “Module Select” stands for module type, including two optional modes, 

named “18*0.5” and “24*0.2”. The Module type will be preset for the type 
of  unit  you  purchased.    If  a  different  size  block  is  later  installed,  this 
setting will need to be changed.  

1.2.3  “Control  Mode”  stands  for  default  control  model  after  power-on, 

including two optional modes, named “Block” and “Tube”. 

 

“Block” mode 

offers basic temperature control during the course of cooling & 

heating only. Compared to the block, the cooling & heating rate of the samples 
in  the  block  are  a  little  slower,  therefore,  the  actual  rate  of  change  of  the 
sample temperature is different from the block temperature. The actual time for 
the sample to reach the block temperature is longer than the time it takes for 
the block to reach the set temperature.  Therefore the dwell time of the sample 
at the set temperature can be shorter than the dwell time of the block at the set 
temperature.(Please  refer  to  the  following  figure:  relation  between  block 
temperature and reagent temperature.) 
 

         
 
 
 
 
 

 

System Parameter Setting 
Interface 

 

Block 

Reagent 

Hotlid=off                                  2009-02-28 10:06:00 

 

Date(yy-mm-dd):

  

09-02-28

 

 

Time(hh-mm-ss):

  

10:06:19

 

       

Module Select:     

24*0.2 

 

 

  

 

Control Mode:      

Block  

 

 

   

     

Hotlid Temp:

        

OFF   

 

 

 

 

Block Temp:         

10C   

  

 

Key Sound:

          

 OFF  

 

 

  

 

Run End Sound:   

 ON   

 

 

 

 

 

                                                  Back         Save  

5

    

 
 
 
 
  

Summary of Contents for MultiGene Mini

Page 1: ...MultiGene Mini Thermal Cycler TC020 24 TC020 24 230V TC050 18 TC050 18 230V Operation Manual Ver 1 2 Labnet International...

Page 2: ...3 1 Description 3 2 Keyboard Diagram 4 3 Key Functions 4 4 Inspection before Power up 5 5 Power on Procedures 5 CHAPTER 3 OPERATION GUIDE 6 1 How to Run a Program 6 1 1 How to edit a program 7 1 2 How...

Page 3: ...0Hz 220 240V 200W 50 60Hz Note Before powering on please confirm that the above operating conditions are satisfied The unit must be plugged into a grounded outlet 2 Transportation and Storage Conditio...

Page 4: ...of cycles 99 Max Segments within a cycle 9 Max temperature time 99 min 59 sec Max incubation and in situ time 99 hour 59 min Program storage capacity 100 5 Software Functions File editing and saving...

Page 5: ...er on 1 Description Heated lid Block LCD Display Keyboard RS232 interface Power switch Power connecter Fuse Socket Note 1 The fuse specifications are as follows 250V 2 5A 125V 5A 5 20mm The fuse shoul...

Page 6: ...g file parameter setting Up Down moving the cursor one position up or down during file parameter setting Enter press it to accept the present parameter settings displayed on the screen Function press...

Page 7: ...e power socket 5 Power on Procedures Turn on the power i e turn the power switch to position 1 the MultiGene Mini Thermal Cycler will beep 4 times and the LCD will display This means the device is sel...

Page 8: ...an be tightened by turning it clockwise After closing the cover but before power on turn the knob clockwise until you hear a click then turn the knob one more turn When the knob is turned clockwise or...

Page 9: ...s below If the current Control Mode is Block mode press Accept button to enter file running interface If the current Control Mode is Tube mode press Accept button a dialog box will be displayed to inp...

Page 10: ...tected press Run a dialog box will be displayed to input the password Press Back to go back the file list interface Press Accept to confirm the password After pressing Run in the file list interface a...

Page 11: ...by pressing Up Down and Right Left Press the keys to modify parameter settings Press CE to change Press Seg to add a segment In every segment you can set temperature Temp duration Time Ramping rate R...

Page 12: ...normally saved File has been saved Please continue another operation will appear Press Enter to close the reminder window If the file name was already in use the system will display Rename or continu...

Page 13: ...ess Stop and Do you want to stop running will be displayed Select F1 or F3 You can then select either Stop or Start Press Pause and the display will show Pause state now Press F3 to continue running Y...

Page 14: ...trol Mode stands for default control model after power on including two optional modes named Block and Tube Block mode offers basic temperature control during the course of cooling heating only Compar...

Page 15: ...temperature which allows the sample to reach the set temperature faster and will help the stay at the set temperature for a longer period Please refer to the following figure relation between block te...

Page 16: ...r bad temperature control accuracy Cooling module failure Contact Labnet or your distributor Cooling module failure 3 Significant change in cooling speed or unavailable temperature below ambient Fan f...

Page 17: ...oken there is a short circuit or the fan is not working 4 Error Message when running Please contact the distributor and Temperature sensor No 1 is disconnected Block sensor No 1 reports an open circui...

Page 18: ...ATER LID J2 4_HEATER CPU PCB4 BYQ5041041000000_CPU J13 1_HV J13 2_HV J13 3_VIN5 J13 4_HOTLID J13 5_FAN J13 6_PWM J13 7_HEAT J13 8_COOL J3_CPU_JTAG_IDC20 J10_KEY IN J5_LCD_24P J9_CPLD_JTAG_IDC10 J16 1_...

Page 19: ...P J9_CPLD_JTAG_IDC10 J16 1_SEN HOTLID J16 2_SEN HOTLID J15 1_SEN FAN J15 2_SEN FAN J15 3_SEN BLOCK J15 4_SEN BLOCK VB1_SUBD9_RS232C J17 1_H J17 2_H L U5 1 2 3 4 5 6 L I N P E N I N N L P E DCFAN U4 1...

Page 20: ...date 9 Time Time HH MM SS 00 00 00 23 59 59 hour minute second 10 Default File The default file name Max length is 11 characters 11 Default User The default user name Max length is 11 characters Block...

Page 21: ...19 Image 2 Image1 block reagent block reagent...

Page 22: ...ts in original material or workmanship This warranty does not cover motor brushes fuses light bulbs batteries or damage to paint or finish Claims for transit damage should be filed with the transporta...

Reviews: