background image

 

 

Note: 

The clinical centrifuge is calibrated at 120 volts line voltage. Operating the 

centrifuge at a line voltage other than 120V can significantly impact the speed settings 
below 2,000 rpm.  

 

4.  Proper loading of the rotor  

4.1  Rotor capacity 

The rotor has been designed to handle samples with a maximum density of 1.2g/mL. 
Tubes should not be run with any substance having a density above 1.2g/mL.  

 

4.2  Rotor loading 

When loading the rotor you should proceed according to sketch A. Fill the tubes equally 
by eye and insert them into the tube shields of the rotor. The difference in weight 
between the tubes should not exceed 2 – 3 grams.  

A partially loaded rotor may be centrifuged. For example, you may operate a 6-place 
rotor with 2, 3 or 4 loaded tubes, but it is important that the occupied tube shields are 
symmetrically arranged. (see sketch A) 

 

 

Sketch A: Examples of properly loaded rotors.  

 

 

 

 

 

 

 

Summary of Contents for ESCZ100A

Page 1: ...Rev 2 August 2016 Z100A Clinical Centrifuge User Manual ESCZ100A...

Page 2: ...About This Manual This manual is designed to assist you in the optimal usage of your Z100A Clinical Centrifuge...

Page 3: ...r is spinning NEVER use solvents or flammables near this or other electrical equipment NEVER centrifuge flammable explosive or corrosive materials NEVER centrifuge hazardous materials outside of a hoo...

Page 4: ...n Centrifuge Installation 3 Operation 3 Closing the lid Lid release Lid lock Starting the centrifuge 4 Proper use of the Rotor 3 Overloading the rotor Balancing the rotor Rotors and accessories 5 Serv...

Page 5: ...rly The centrifuge is partially dismantled Service has been attempted by unauthorized or unqualified personnel The rotor has not been installed securely on the motor shaft Rotors and accessories not b...

Page 6: ...ks until it has been established that your unit is working properly The accessories supplied with the centrifuge should be kept with the instruction manual near the centrifuge s place of installation...

Page 7: ...ge run has been completed 3 3 Lid lock The centrifuge can only be started with the lid closed Do not attempt to open the lid until the rotor comes to a complete stop 3 4 Starting the centrifuge To sta...

Page 8: ...run with any substance having a density above 1 2g mL 4 2 Rotor loading When loading the rotor you should proceed according to sketch A Fill the tubes equally by eye and insert them into the tube shie...

Page 9: ...plastic After cleaning ensure that all parts are dried thoroughly by hand or in a warm air cabinet max temp 50 C 5 3 Disinfection The centrifuge should be disinfected periodically and before shipping...

Page 10: ...lace if necessary see 5 5 Lid lock will not release Possible reason Lid lock is jammed Solution Push front of lid down while pushing release button Centrifuge cannot be started although power is on Po...

Page 11: ......

Page 12: ...lectronic equipment WEEE Labnet Z100A Clinical Centrifuge is marked with the crossed out wheeled bin and must not be disposed of with domestic waste Consequently the buyer shall follow the instruction...

Reviews: