AccuBlock Digital Dry Bath | 1
1.0 Safety Precautions
Before using the
AccuBlock
Digital Dry Bath for the first time, please read this entire manual carefully.
To guarantee problem free, safe operation, it is essential to observe the following points:,
1.1 Operation Safety Precautions
•
Do not
use this product in an explosive environment
•
Do not
use in the presence of flammable or combustible material
•
Do not
heat substances that react violently when heated
•
Do not
touch block when hot or when unit is heating. Use block lifter.
•
Do not spill liquids into the well area or into the unit side vent holes
•
Connect unit only to a properly grounded outlet
2.0 General Specifications
Warning!
Do not touch area around block or block well when unit is hot. This can result
in bodily injury, including burns.
Warning!
Modification of this product or use of the product in any manner not specified
by this instruction manual may cause injury and/or may void the warranty.
If the equipment is used in a manner not specified by the manufacturer, the
protection provided by the equipment may be impaired.
Temperature range
A5°C to 150°C
Temperature display resolution
0.1°C 4 digit LED
Temperature uniformity
± 0.2°C (at 37°C in block)
Temperature accuracy
± 0.3°C
Temperature controller
Microprocessor – user calibratable
Timer
1 to 99hours59minutes or continuous in 1 minute increments
Operating Altitude
Not to exceed 2,000 meters
Control
PI Microprocessor controller
I/O
USB unidirectional
Dimensions (W x D x H)
D1301/D1302 8.3 x 11.4 x 4.7 in./21 x29 x 12cm
D1304 8.3 x 15.3 x4.7 in./21 x 39 x 12 cm
Weight
D1301/D1302 7.0 lbs./3.2 kg
D1304 9.6 lbs./4.4 kg
2.0 Electrical requirements:
D1301
115V
50/60Hz, 1 A, Fuse 1.6 AT
D1301-230V
230V
50/60Hz, 0.5 A, Fuse 1AT
D1302
115V
50/60Hz, 1.9A, Fuse 3.15AT
D1302-230V
230V
50/60Hz, 0.95 A, Fuse 1.6AT
D1304
115V
50/60Hz, 3.73 A, Fuse 5 AT
D1304-230V
230V
50/60Hz, 1.86 A, Fuse 3.15 AT
Summary of Contents for Accublock D1301