7 | Basic Operations
This section describes Access system basic operations, including:
l
l
l
l
l
Shutting Down The Access System
l
Recovering From An Emergency Shutdown
Starting The Access System
1. Check the following items before starting the Access system:
a. House air is supplied.
b. Vacuum air is supplied.
c. No E-Stops are triggered.
d. Main Disconnect Switch is turned on.
e. System Reset button is solid blue.
f. Teach key is turned to Run Mode.
2. Turn on each device. Refer the specific device user guide for instructions.
3. Launch Tempo software and navigate to
Control Panel > Alarms
to view the state of the Access system.
The
Safety
,
System Resources
, and
Environment
alarms need to be green indicating that the system is in an
idle state and ready to start normally. The
Run Mode
alarm needs to be white indicating that the system is in
Run Mode. The
Building Management System
alarm indicators need to be gray indicating that the system is
running on electric power (as opposed to running on generator or battery power).