Figure 53: Echo Standalone Mode
Figure 54: Echo Standalone Position - Abort Run Dialog Box
7. The run, which the Echo Turntable is reserved for, is paused and the dialog box above lists two option:.
a. To continue a run, you must reset the Standalone mode sensor by returning the Echo Turntable locked back into
home position, and then the protocol can be continued.
b. To abort the protocol run, click the Abort Run button to enable access to transition the Echo Turntable to
Standalone mode.
c.
Note:
A manual transition to or from
Standalone
mode is not allowed during a protocol run. The same
applies to
Online
and
Offline
modes, which are also not allowed to be changed during a protocol run.
8. Close the Tunnel door.
9. Verify that the Echo Turntable Standalone position and Tunnel door Closed mode are as shown below.
10. Verify that the Echo Turntable is in Standalone mode by checking the device status in the device drop-down menu in the
Tempo Hardware tab.
11. Verify that the Echo Turntable is in Standalone mode by checking the device status in the device drop-down menu in the
Tempo Hardware tab.
56
PN | 001-15357
USER GUIDE | Access™ Dual Robot System
Using The Echo Turntable For Standalone Use