11
2: Before You Install
Before you install the product, the installation site should be planned. Examine the
location where you intend to install it. You must be certain that the area is level and of
solid construction. In addition, a hot water source, a drain, a purified water source (if
applicable), and an electrical supply must be located near the installation site. A cold
water supply is required when the optional Cool Drain Water Kit is installed.
Location Requirements
The Glassware washer should be located in an area that is in close proximity to the
required utilities.
Clearance Requirements
If your glassware washer is to be mounted undercounter model, you must ensure the
undercounter opening is the proper size to accommodate the washer. The location
should provide adequate clearance for the door to lay down with access to load and
unload the racks.
See
Appendix B: Dimensions
for overall product dimensions.
Electrical Requirements
The product models have the following electrical requirements.
Catalog Number
Heater
Watts
Typical Operating
Current (Amps)
Electrical Circuit Requirements
1
4xxx01000
1.2kW
16 A
115 V, 60 Hz, 20 A
1 Phase
4xxx01010
2.0kW
11 A
208-230 V, 50/60 Hz, 20 A
1 Phase
4xxx01013
6.0kW
21/19 A
2
208/230/400 V, 50/60 Hz,
3 Phase
1 Electrical Requirements, ‘V’ = VAC (Voltage with alternating current), ‘A’ = Amperes
2
Operating with 208V, current draw approximately 21 A; Operating with 230V, current drawing approximately 19 A.
Summary of Contents for SteamScrubber 4010010 Series
Page 101: ...101...