Fuse Replacement
(1)
Unplug the unit from electrical service.
(2)
Using a small screwdriver or coin, turn the slotted fuse holder ¼ turn
counterclockwise.
(3)
Pull the fuse and holder straight out. Remove the fuse from its holder
by pulling the fuse straight out.
Replace the fuse with a proper replacement part only. DO NOT
attempt to replace the fuse with one of a higher amperage rating.
(4)
To reinstall the fuse, reverse the above steps.
Removing the Boiler
(1)
Unplug the RapidStill II from electrical service.
(2)
Turn the unit around to gain access to the boiler flask.
(3)
Remove the head of the boiler by pulling it straight up. If the head is
stuck onto the boiler, apply water to the joint to loosen it.
(4)
Pull the ring that secures the boiler straight up until it clears the neck of
the boiler flask. The boiler can now be removed.
(5)