Location
The RapidStill II should be located on a level work surface, adjacent to
appropriate electrical, water and drain connections.
Electrical Connection
The electrical outlet should be rated for 115 VAC, 60 Hz, 15 Amps, and dedicated
to the RapidStill II.
Water Connection
The feed water should be municipal drinking water grade and of sufficient quality
to prevent or minimize scaling of the boiler. The feed water should have a
minimum supply volume of 1 gallon (3.8 liters) per minute, at 10 PSI.
Using the supplied braided water line, connect the tube to the water inlet fitting on
the rear of the RapidStill II as shown in Figure 3. To connect or disconnect the
fittings, follow the instructions in Figure 4, below. If your feed water promotes
excessive scaling of the boiler, or has high NH
3
levels, you may wish to install the
Optional Water Inlet Kit, part #6544200.