Chapter 3: Getting Started
Product Service 1-800-522-7658
11
Original instructions
Good Installation
– Picture 1
Good Installation
– Picture 2
Poor Installation
– Picture 3
Two-piece C-clamp coupler supports
exhaust duct diameter to secure weight load
of duct picked up by structural support.
Structural support is connected to C-clamp coupler.
Note:
Duct diameters, duct materials,
and heights change and Labconco
does not carry the structural supports
or guy wires.
Guy wires from 3 sides are supported
off the roof to prevent wind velocities
from deflecting the exhaust duct.
Blower damage can occur without
structural supports and guy wires.
(Please note the crooked exhaust duct
when installed without structural
supports and without guy wires.)