Chapter 3: Getting Started
Product Service 1-800-522-7658
8
Original instructions
IF YOUR BLOWER WAS DAMAGED IN TRANSIT, YOU
MUST FILE A CLAIM DIRECTLY WITH THE FREIGHT
CARRIER. LABCONCO CORPORATION AND ITS
DEALERS ARE NOT RESPONSIBLE FOR SHIPPING
DAMAGE.
Do not discard the packing material for your blower until you have installed and
tested the blower.
Install the Blower on a Supporting
Structure
Now that you have located your blower as instructed in
Chapter
2: Prerequisites
,
you are ready to mount the blower on a roof curb support. Additionally, you may
isolate the blower by mounting the blower on either vibration isolators or vibration
mounting pads, which are available from a local industrial supply company. The
supporting structure is custom for each installation. Labconco recommends
supporting the blower with 5/16" diameter mounting hardware. See
Appendix B:
Blower Dimensions
for appropriate mounting hole locations for your particular
blower.
Adjust the Blower Outlet Orientation
The housing on your blower can be rotated to facilitate your duct run
configuration. By rotating the blower housing, you will be able to change the
blower outlet location on your blower. The Coated Steel Blowers can be rotated to
one of eight different positions and the Fiberglass or PVC Blowers can be rotated
to one of three different positions. See the drawings in
Appendix A
to locate the
parts to disassemble. To rotate the blower housing, proceed as follows:
1.
Remove the upper weathercover panel, which has louvers.
2.
Once removed, you have complete access to the 8 fasteners that hold your
blower housing assembly in position. Removing these fasteners allows you
to rotate the housing to one of the other desired outlet positions. Then
insert the 8 fasteners and secure the blower housing in that specific
position.
3.
Reinstall the upper weathercover panel and you are ready for operation.
HINT:
Most blowers are installed for upward discharge.
!
Summary of Contents for Intelli-Sense CS 10
Page 36: ...Chapter 5 Using Your Blower Product Service 1 800 522 7658 32 Original instructions...
Page 48: ...Chapter 7 Modifying Your Blower Product Service 1 800 522 7658 44 Original instructions...
Page 58: ...Chapter 9 Troubleshooting Product Service 1 800 522 7658 54 Original instructions...
Page 62: ...Product Service 1 800 522 7658 58 APPENDIX B BLOWER DIMENSIONS...
Page 63: ...Appendix B Blower Dimensions Product Service 1 800 522 7658 59 Original instructions...
Page 64: ...Appendix B Blower Dimensions Product Service 1 800 522 7658 60 Original instructions...
Page 72: ...Product Service 1 800 522 7658 68 APPENDIX F TYPICAL PERCHLORIC ACID HOOD INSTALLATION DIAGRAM...