96
flow must be at least
1.25 gallons (4.7
liters)/minute.
Improper water
supply pure.
Pure Fill All The
Way tests Pure
fill time and flow
rate.
Fail: Make sure pure
water is turned on
with no restrictions.
DI water supply
must have volume
of 3.4 gallons (4.7
liters) per fill and
flow of at least 0.9
gallons (3.4
liter)/minute.
Defective fill valves.
Fill Valve Test
(Hot Top or
Pure): Test valve
operation.
Pass: Clean screens
inside the inlet
fittings of the water
valves.
Fill Valve Test
(Hot Top or
Pure): Test valve
operation.
Fail: If valve is
getting power
replace valve. If no
power, check wiring
or IOB output
Defective DI water
pump.
Pure Pump Test:
Test pump
operation.
Fail: If pump is
getting power
replace. If no power,
check wiring or IOB
output
Are leak in hose
connections to water
level sensor.
Check Sensor
Screen: >75mm
when full.
Fail: Repair air leak.
Defective fill level
sensor.
Check Sensor
Screen: <75mm
when full.
Pass: Recalibrate
sensor. Contact
Labconco Product
Service.
Defective controller.
Check output from
IOB to both valves
and DI pump.
Fail: Replace IOB.
Firmware
FI
Firmware not
responding.
Power Reset,
reinstall firmware or
Factory Reset
High Water
HW
Water level set point
to high.
FILL ALL THE
WAY Water level
over distribution
tube.
Pass: Recalibrate
sensor. Contact
Labconco Product
Service.
Inlet Valve failure.
Check water inlet
in side of tank for
water leaking
when unit off.
Replace Inlet Valve