Chapter 7: Troubleshooting
Product Service 1-800-522-7658
28
Other corrective actions for potential problems are as follows:
PROBLEM
CAUSE
CORRECTIVE ACTION
Unit will not operate
Unit not connected to electrical
power
Connect unit to proper electrical receptacle.
Circuit breaker blown
Correct electrical problem and reset circuit
breaker by pressing button.
Lid open
Close lid.
Excessive vibration
Sample tubes not located
symmetrically in rotor
Reposition sample tubes.
Sample odor in lab
Vent hose exhausting into lab
area
Redirect hose to fume hood.
Evaporation rate is reduced
Heater inoperable
Contact Labconco.
Vacuum pump failure
Check pump.
Obstruction in hose
Remove obstruction or replace hose.
Lack of adequate vacuum
See below.
No vacuum/poor vacuum
Pump not on
Turn on pump.
Control valve open
Check control valve.
Leaks in lines or connectors or
gasket
Locate and repair.
Foreign material on lid gasket
Clean gasket and lid.
Pump is not functioning properly
Check pump by locating vacuum gauge closer to
pump and close off rest of system. Check pump
oil for cloudiness or particles and change.
If pump is faulty, seek authorized service or
replace pump.
Cold Trap or Concentrator
gasket is not sealing properly
Check gasket for cleanliness. Adjust gasket as
needed.
Cold Trap lid not seated
Hold lid down until vacuum is initiated.
Ice formed on Cold Trap lid
sealing surface
Defrost and wipe dry.
Secondary Trap Canister not
fitted properly
Tighten all connections to and from the
Secondary Trap Canister.
Secondary Trap Insert is spent
Replace with new insert.
New Secondary Trap has
moisture in it
Run vacuum pump for 24 hours to remove
moisture.
Summary of Contents for CentriVap 73100 Series
Page 1: ...CentriVap Refrigerated Centrifugal Concentrators Models 73100 Series...
Page 31: ...Chapter 7 Troubleshooting Product Service 1 800 522 7658 27...
Page 35: ...Appendix A CentriVap Refrigerated Concentrator Components Product Service 1 800 522 7658 31...
Page 36: ...Appendix A CentriVap Components Product Service 1 800 522 7658 32...
Page 37: ...Product Service 1 800 522 7658 33 Appendix B CentriVap Refrigerated Concentrator Dimensions...