Chapter 6: Maintaining the Ductless Enclosure
Product Service 1-800-522-7658
49
Organic Sensor Board Replacement
If the organic sensor board is replaced, the Paramount must be calibrated for
the “Organic Sensor.” Contact Labconco Product Service. See
Appendix A:
Replacement Parts
and reference Figure 6-5. Please note that the 6'
Paramount utilizes two organic sensor boards, one for each motorized
impeller.
1.
Consult the wiring diagram on the back of the front panel and disconnect
all power.
2.
Remove the organic sensor connector as shown in Figure 6-5.
3.
Remove the four Phillips screws that mount the organic sensor board.
4.
Install the new organic sensor board in reverse order.
Note:
Be careful
with the delicate electronic sensors on the organic sensor board and do
not move them or rotate them.
5.
Follow the “Calibration” menu screens from
Chapter 5
as the Organic
Sensor needs calibration. A separate code and procedure is required and
Product Service must be contacted.
Figure 6-5
Organic Sensor Board