Remote control
7-13
SMAN
This setting command puts the ER8000 in manual mode. The reply is always
MAN=OK
which acknowledges the command has been granted.
The ER8000 goes to manual mode.
Example:
#SMAN *
SMANP
This setting command puts the ER8000 in manual mode and sets the default value
to the following settings:
•
RBW Automatic
•
Attenuator Automatic
•
MinAtt=10 dB
•
Preselector On
•
Preamplifier OFF
•
Hold Time =1000ms
•
Pulse Limiter = OFF
The reply is always
MANP=OK
which acknowledges the command has been
granted.
The ER8000 goes to manual mode.
Example:
#SMANP*
SMAT a
This setting command sets Attenuator, for Manual Mode, to the value indicated by
(a) which can be 0 to 45. Automatic Attenuator feature is disabled. If (a) is a negative
figure then the Automatic Attenuator feature is turned on. The reply is
MAT=OK
which acknowledges the command has been granted or
MAT =SERR
if the
command has been ignored.
When the command is granted the ER8000 goes to manual mode automatically.
Example:
#SMAT 15*
(set attenuator to 15dB and disable Automatic Attenuator
feature)
Example:
#SMAT -1*
(Enable Automatic Attenuator feature)
SMHT h
This setting command sets hold time for manual mode. The string (h) should be
expressed in ms. The reply is
MHT=OK
which acknowledges the command has
been granted or
MHT =SERR
if the command has been ignored.
When the command is granted the ER8000 goes to manual mode automatically.
Example:
#SMHT 1500*
Sets the hold time, for manual mode, to 1500 ms (or 1.5s)
SRBW b
This setting command sets RBW. The argument (b) should be a string representing
the RBW as follows:
o
0
RBW is Automatic
so that the ER8000 automatically selects
the suitable RBW according to the tuned frequency
o
1
300kHz
o
2
100kHz
o
3
30kHz
o
4
10kHz
o
5
3kHz
o
6
9kHz (Cispr 16-1)
o
7
200Hz (Cispr 16-1)
o
8
1kHz
o
Only for C, D band o > 30 MHz
10
120kHz (Cispr 16-1) (Opt. 01)
9
1MHz (Opt. 01)
The reply is
RBW =OK
which acknowledges the command has been granted or
RBW =SERR
if the command has been ignored.
Example:
#SRBW 3*
sets the RBW to 30kHz and disables Auto RBW mode
Example:
#SRBW 0*
turn the RBW mode in Automatic
Summary of Contents for NARDA PMM ER8000
Page 28: ...2 10 Installation This page has been left blank intentionally ...
Page 38: ...4 4 Applications Fig 4 2 ER8000 internal LISN Fig 4 3 ER8000 with external L2 16B LISN ...
Page 40: ...4 6 Applications This page has been left blank intentionally ...
Page 52: ...6 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally ...
Page 82: ...Suggerimenti Commenti Note Suggestions Comments Note ...