1.7 Main Specifications
Table 1-1 lists the PMM ER8000 performance specifications.
The following conditions apply to all specifications:
•
The ambient temperature shall be -5°C to 45°C
TABLE 1-1 Main Specifications
Frequency
Range
Resolution
Reference frequency
9 kHz to 30 MHz (Opt. 00)
9 kHz to 3 GHz (Opt. 01)
1 Hz; 100Hz above 30 MHz (Opt. 01)
< 2.5 ppm
Spectrum analysis method
FFT, size up to 8192, minimum overlap 89%
RF input
VSWR
10 dB RF att.
0 dB RF att.
Attenuator
Preamplifier gain
Pulse limiter
Z
in
50 Ω,
N fem.
< 1.2; < 2 above 1 GHz
< 2
0 dB to 45 dB (5 dB steps)
20 dB; 10 dB above 30 MHz Low saturation preamplifier (after preselector)
Built in (selectable) below 30 MHz
Max input level
(without equipment damage)
Sinewave AC
Voltage pulse spectral density
Max. pulse voltage
Max. DC voltage
140 dBuV (2 W); 137 dBuV (1 W) above 30 MHz
176 dBuV/MHz below 150 kHz; 130 dBuV/MHz below 30 MHz; 97 dBuV/MHz below 1
GHz
200V (≤ 20 us)
50V
Preselector
Frequency ranges
(Seven BP filters - 15 MHz BW to ADC)
9 kHz to 150 kHz
30 MHz to 96.6 MHz tracking (Opt. 01)
150 kHz to 15 MHz
96.6 MHz to 311 MHz tracking (Opt. 01)
15 MHz to 30 MHz
311 MHz to 1000 MHz tracking (Opt. 01)
1 GHz to 3 GHz (Opt. 01)
IF bandwidth
6dB
CISPR 16-1-1
100 Hz, 300 Hz
1 kHz, 3 kHz, 10 kHz, 30 kHz, 100 kHz, 300 kHz
1 MHz, 3 MHz
200 Hz, 9 kHz, 120kHz, 1 MHz
Displayed Average Noise Level
Preselector OFF, preamplifier OFF, Ht 1s
Preselector ON, preamplifier OFF, Ht 1s
Preselector ON, preamplifier ON, Ht 1s
9 kHz to 150 kHz (200 Hz RBW)
< -17 dBuV
0.15 MHz to 30 MHz (9 kHz RBW)
< 0 dBuV
30 MHz to 300 MHz (120 kHz RBW)
< 4 dBuV (Opt. 01)
300 MHz to 3 GHz (120 kHz RBW)
< 10 dBuV (Opt. 01)
9 kHz to 150 kHz (200 Hz RBW)
< -14 dBuV
0.15 MHz to 30 MHz (9 kHz RBW)
< 3 dBuV
30 MHz to 300 MHz (120 kHz RBW)
< 1 dBuV (Opt. 01)
300 MHz to 3 GHz (120 kHz RBW)
< 6 dBuV (Opt. 01)
9 kHz to 150 kHz (200 Hz RBW)
< -27 dBuV
0.15 MHz to 30 MHz (9 kHz RBW)
< -14 dBuV
30 MHz to 300 MHz (120 kHz RBW)
< -5 dBuV (Opt. 01)
300 MHz to 3 GHz (120 kHz RBW)
< 0 dBuV (Opt. 01)
Detectors
Peak, Quasi-Peak, Average, RMS, RMS-Average, C-Average.
Smart Detector function above 30 MHz (Opt. 01)
Summary of Contents for NARDA PMM ER8000
Page 28: ...2 10 Installation This page has been left blank intentionally ...
Page 38: ...4 4 Applications Fig 4 2 ER8000 internal LISN Fig 4 3 ER8000 with external L2 16B LISN ...
Page 40: ...4 6 Applications This page has been left blank intentionally ...
Page 52: ...6 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally ...
Page 82: ...Suggerimenti Commenti Note Suggestions Comments Note ...