Possible number of tests with fresh batteries.
Continuity :Approx. 2000 times min. at load 1Ω
Insulation Resistance :Approx. 1000 times min. at load 1MΩ (1000V)
LOOP/PFC/PSC :Approx. 1000 times min. (ATT)
RCD :Approx. 2000 times min. (G-AC X1 30mA)
EARTH :Approx. 1000 times min. at load 10Ω
VOLT/PHASE ROTATION :Approx. 50H
Reference Conditions
Ambient temperature
23±5
o
C
Relative humidity
45% to 75%
Nominal system voltage and
frequency
230V, 50Hz
Altitude
Less than 2000m
5.2 Operating error
Continuity (EN61557-4)
Operating range compliant with
EN61557-4 operating error
Maximum percentage
operating error
0.20~1999MΩ
±30%
The influencing variations used for calculating the operating error are denoted as follows;
Temperature : 0
o
C and 35
o
C
Supply voltage : 8V to 13.8V
Insulation Resistance(EN61557-2)
Volt
Operating range compliant with
EN61557-2 operating error
Maximum percentage
operating error
250V
0.25~199.9MΩ
500V
0.50~1999MΩ
1000V
1.00~1999MΩ
±30%
The influencing variations used for calculating the operating error are denoted as follows;
Temperature : 0
o
C and 35
o
C
Supply voltage : 8V to 13.8V