5
—
3.Specification
Applicable standards
IEC 61557-1,2,4
IEC 61010-1, -2-030 Masurement CAT III 600V Pollution degree2
(Location for use: altitude 2000m or less, indoor use)
IEC 61010-031
IEC 60529
IP40
EN 61326-1 (EMC)
EN 50581 (RoHS)
Measuring range and accuracy
<Insulation resistance range>
KEW3021A
Nominal voltage
1000V
500V
250V
125V
Auto Range
0
〜
4/40/400/2000M
Ω
0
〜
4/40/200M
Ω
Open circuit
voltage
Nominal test voltage 100%~120%
Short circuit
current
Within 1.5mA
Nominal test
current
1M
Ω
0.5M
Ω
0.25M
Ω
0.125M
Ω
1mA 0%
〜+
20%
1
st
effective
measurement
range
0.200
〜
1000M
Ω
0.200
〜
200.0M
Ω
0.200
〜
40.00M
Ω
0.200
〜
20.00M
Ω
Center scale
range
50M
Ω
50M
Ω
5M
Ω
5M
Ω
Accuracy
±
(2%rdg
+
6dgt)
2
nd
effective
measurement
range
0.110
〜
0.199M
Ω
1001
〜
2000M
Ω
200.1
〜
2000M
Ω
40.01
〜
2000M
Ω
20.01
〜
200.0M
Ω
Accuracy
±
(5%rdg
+
6dgt)
Accuracy at 0
Within 6dgt
KEW3022A
Nominal voltage
500V
250V
100V
50V
Auto Range
0
〜
4/40/400/2000M
Ω
0
〜
4/40/200M
Ω
Open circuit
voltage
Nominal test voltage 100%~120%
Short circuit
current
Within 1.5mA
Nominal test
current
0.5M
Ω
0.25M
Ω
0.1M
Ω
0.05M
Ω
1mA 0%
〜+
20%
1
st
effective
measurement
range
0.200
〜
200.0M
Ω
0.200
〜
40.00M
Ω
0.200
〜
20.00M
Ω
0.200
〜
20.00M
Ω
Center scale
range
50M
Ω
5M
Ω
5M
Ω
5M
Ω
Accuracy
±
(2%rdg
+
6dgt)
2
nd
effective
measurement
range
0.110
〜
0.199M
Ω
200.1
〜
2000M
Ω
40.01
〜
2000M
Ω
20.01
〜
200.0M
Ω
20.01
〜
200.0M
Ω
Accuracy
±
(5%rdg
+
6dgt)
Accuracy at 0
Within 6dgt
www.
GlobalTestSupply
.com