18
—
number displayed at the lower left flickers.Select any memory
number with the UP or DOWN key, and press the
key.
(2) Then, the measurement result displayed at the middle on the LCD
flickers. Press the
key to save the result.It will be saved to
the internal memory, and the memory number is increased by 1.
Press the
key to delete the result.
Recalling the memory data
(1) Set the Range selector switch to any desired Insulation resistance
range, and press the
key.
(2) Press the UP or DOWN key to switch the memory number.The
measured voltage and value are displayed on the LCD.
Deleting all memory data
(1) To delete all memory, rotate the Range selector switch from OFF
to
Ω
(CONTINUITY) position with the
key pressed down.Then
release the
key.
(2) The message "clr" is displayed on the LCD. It flickers and buzzer
sounds when the
key is being pressed down again. The
deletion is completed once "
Ω
" is displayed on the LCD.
Note:
The mark "---" is displayed on the LCD when no data is saved with
the selected memory No.
Resistance value is displayed on the LCD when data is saved with
the selected memory No.
Old data is overwritten with new when a new data is saved with the
same memory No.
8.Fuse & Battery replacement
#
DANGER
Never open the Battery cover during a measurement.Dispose the used
batteries according to the rules, which is defined by each community.
#
WARNING
To avoid possible electric shock, remove test leads before opening
the Battery cover. After replacing batteries, be sure to tighten up the
screws for Battery cover.
#
CAUTION
Do not mix new and old batteries.
Install batteries in correct polarity as marked inside the Battery
compartment.
8-1 Battery Replacement
(1) Disconnect the test probe from the instrument.
(2) Open the battery compartment cover by unscrewing the metal
captive screw to reveal battery compartment.