̶ 7 ̶
●Applicable
standards
●IEC 61010-1, 2-030 CAT III 600V
Pollution degree 2
●IEC 61557-1, -2
●IEC 60529 IP40
●IEC 61326-1, 2-2
●IEC 61010-031
MODEL7260…CAT III 600V (w/ cap)
CAT II 1000V (w/o cap)
CAT II 1000V (w/ 8017A)
( Attach the supplied protective cap to use this
test leads in CAT III or higher environments.)
MODEL7261A…CAT III 600V (w/ alligator clip)
CAT II 600V (w/ flat test bar)
( Attach the alligator clip to use this test leads
in CAT III or higher environments.)
※ When test probes, sometimes with metal tips,
are connected and used with the instrument,
the measurement category and voltage rating
of the lowest rated item is applied.
●Location for use Altitude 2000m or less, in-door use
●Operating temp.
& humidity range 0ºC to 40ºC, 80% or less (no condensation)
●Storage temp.
& humidity range -10ºC to 50ºC, 75% or less (no condensation)
●Withstand
voltage
5160V AC (50/60Hz) / 5 sec
Between electrical circuit and enclosure
●Insulation
resistance
50MΩ or more/ 1000V DC
Between electrical circuit and enclosure
●Auto-power-off
Turns off the instrument automatically, after a
beep sound, if there is no function change, range
change or button press for about 10 min.
(※not work during a measurement)
●Scale light/
LED light
Automatically turns off if there is no activity for
about 2 min. (※disabled during a measurement)
●Dimensions
97 (L) × 156 (W) × 46 (H) mm
●Weight
Approx. 430g (including batteries)
●Power source
Four size AA batteries
※Use of alkaline battery is recommended.