be issued immediately as soon as the meter is in “wait for event” state.
How to use Immediate Trigger
Use the following command in your PC terminal to set the internal
immediate trigger.
TRIGger:SOURce IMMediate
C. Single Trigger Mode (Front panel operation only)
Definition
Single trigger mode takes one reading (or specified number of readings)
each time when user presses TRIGGER key. (Please refer to 4.2.3 for
setting the number of samples on each trigger.) When the TRIG
annunciator on the display is lit, the meter is ready for next trigger event.
The single trigger mode is only available through the front panel
operations,
but it’s not available for the diode and continuity functions.
How to use Single Trigger
1.
Press one of the measurement function keys to select a
measurement.
2.
Press TRIGGER key on the front panel to set the trigger mode.
3.
Disable the mode by pressing SHIFT + TRIGGER.
4.2.2 Trigger Source
In M3510A/11A, you can specify the trigger source to be one of these
three options: front panel operations, external hardware trigger source
and remote interface operations.
Front Panel Trigger Operation
Use SHIFT + TRIGGER buttons for auto triggering and TRIGGER button
for single triggering.
External Hardware Trigger
You apply a trigger pulse to the Ext TRIG terminal on the rear panel.
Summary of Contents for M3500
Page 6: ......
Page 10: ......
Page 12: ...w a r ...
Page 13: ......
Page 14: ......
Page 16: ......
Page 29: ......
Page 37: ......
Page 41: ......
Page 42: ...In Short Condition ...
Page 43: ......
Page 44: ......
Page 109: ...104 104 104 104 651 Panel Program Checksum Error The checksum from the panel is incorrect ...
Page 110: ...K U S A M ME C O ...
Page 148: ...K U S A M M E C O K U S A M M E C O ...
Page 155: ...K U S A M M E C O ...
Page 163: ...158 158 158 158 m_nCount 0 m_defaultRM_usbtmc 0 FreeLibrary hUSBTMCLIB hUSBTMCLIB NULL return ...
Page 164: ......