39
When adjusting resistance source, the exciting current is (+) for adjustment point ‘’0’’ and ‘’F’’, and is (-)
for adjustment point “–0” and “–F”.
Turn on the meter; press the
〔
SOURCE
SOURCE
SOURCE
SOURCE
〕
key while simultaneously holding down the
(((
(
MEASURE)
MEASURE)
MEASURE)
MEASURE)
key enters
the source calibration state. LCD shows “CA-0” symbol on the assistance districts part, the present calibrating
point on the main districts part and the high 5 digits of the responding value and its unit. The digit in the right
on the assistance districts part is the lowest digit of the value.
Tips:
If the battery level is below 25% full, the adjustment operation can't be operated. And the LCD shows "ERR"
in the lower part.
12.1
12.1
12.1
12.1
Adjusting
Adjusting
Adjusting
Adjusting
Voltage
Voltage
Voltage
Voltage
Source
Source
Source
Source
Figure 14
Adjusting voltage source
Step
Step
Step
Step
1
1
1
1
:::
:
Using the
〔
mV
mV
mV
mV
〕
key, select DC voltage function. Connect the lead cables for measurement to the
standard digital meter as shown in Figure 14.
30V
MAX
COM
COM
COM
COM
COM
COM
COM
COM
4W
4W
4W
4W
mV
mV
mV
mV
Tc
Tc
Tc
Tc
MEASURE
MEASURE
MEASURE
MEASURE
SOURCE
SOURCE
SOURCE
SOURCE
3W
3W
3W
3W
RTD
RTD
RTD
RTD
PEAK
30V
MAX
PEAK
mV
mV
mV
mV
RTD
RTD
RTD
RTD
Tc
Tc
Tc
Tc
Hi
Lo
数字表(
KEITHLEY
2000)