Kusam-meco KM 805s Operation Manual Download Page 12

®

®

19

MUMBAI

TEST CERTIFICATE

AUTORANGING 

DIGITAL MULTIMETER 

This Test Certificate warrantees that the product has been 

inspected  and  tested  in  accordance  with  the  published 

specifications.

The  instrument  has  been  calibrated  by  using  equipment 

which has already been calibrated to standards traceable 

to national standards.

MODEL NO. ________________

KM 805s

SERIAL NO. ___________

DATE: ___________

ISO 9001

REGISTERED

QC

PASS

KUSAM-MECO

20

WARRANTY

Each “KUSAM-MECO” product is warranted to be free from defects in 
material and workmanship under normal use & service.  The warranty 
period is one year (12 months) and begins from the date of despatch of 
goods.   In case any defect occurs in functioning of the instrument, 
under proper use, within the warranty period, the same will be rectified 
by us free of charges, provided the to and fro freight charges are borne 
by you.
This warranty extends only to the original buyer or end-user customer 
of a “KUSAM-MECO” authorized dealer.
This warranty does not apply for damaged Ic’s, fuses, burnt PCB's, 
disposable batteries, carrying case, test leads, or to any product which 
in “KUSAM-MECO’s” opinion, has been misused, altered, neglected, 
contaminated  or  damaged  by  accident  or  abnormal  conditions  of 
operation or handling.
“KUSAM-MECO” authorized dealer shall extend this warranty on new 
and unused products to end-user customers only but have no authority 
to  extend  a  greater  or  different  warranty  on  behalf  of  “KUSAM-
MECO”.
“KUSAM-MECO’s”  warranty  obligation  is  limited,  at  option,  free  of 
charge repair, or replacement of a defective product which is returned 
to a “KUSAM-MECO” authorized service center within the warranty 
period.
THIS  WARRANTY  IS  BUYER’S  SOLE  AND  EXCLUSIVE 
REMEDY  AND  IS  IN  LIEU  OF  ALL  OTHER  WARRANTIES, 
EXPRESS  OR  IMPLIED,  INCLUDING  BUT  NOT  LIMITED  TO 
ANY  IMPLIED  WARRANTY  OF  MERCHANTABILITY  OR 
FITNESS  FOR  A  PARTICULAR  PURPOSE.    “KUSAM-MECO” 
SHALL  NOT  BE  LIABLE  FOR  ANY  SPECIAL,  INDIRECT, 
INCIDENTAL  OR  CONSEQUENTIAL  DAMAGES  OR  LOSSES, 
INCLUDING  LOSS  OF  DATA,  ARISING  FROM  ANY  CAUSE 
WHATSOEVER.
All transaction are subject to Mumbai Jurisdiction.

Summary of Contents for KM 805s

Page 1: ...h Resistance Tester Digital Panel Meters DC Power Supplies High Voltage Detector Calibrators Gas Analysers Frequency Counter Function Generator Phasing Sticks Battery Tester Waterproof Pen Testers Solar Power Meter EMF Detector Wood Paper Grain Moisture Meter Transistorised Electronic Analog Digital Insulation Resistance Testers upto 10 KV Digital Sound Level Meter Sound Level Calibrator Digital c...

Page 2: ...tal Multimeter TABLE OF CONTENTS TITLE PAGE NO I Safety II Cenelec Directives III Special Features IV General Specifications V Electrical Specifications VI Product Description VII Operation VIII Maintenance IX Test Certificate X Warranty 01 03 03 03 06 10 11 17 19 20 ...

Page 3: ...nstallation Examples are measurements on household appliances portable tools and similar equipment 02 Terms in this manual WARNING Identifies conditions and actions that could result in serious injury or even death to the user CAUTION Identifies conditions and actions that could cause damage or malfunction in the instrument WARNING To reduce the risk of fire or electric shock do not expose this pr...

Page 4: ... Operating below 2000m Pollution degree 2 IV GENERAL SPECIFICATIONS 04 o o Storage Temperature 20 C to 60 C 80 R H With battery removed Temperature Coefficient nominal 0 15 x specified accuracy C o o o o 0 C 18 C or 28 C 40 C or otherwise specified Low Battery Below approx 2 4V Power supply Standard 1 5V AAA Battery x 2 Power Consumption 3 2mA typical Sleep mode Timing Idle for 30 minutes Sleep mo...

Page 5: ... for 400 0 mV range 0 3 rdg 4dgts 0 5 rdg 3dgts 0 5 rdg 3dgts 0 5 rdg 3dgts Accuracy Resolution Range 1000 V 1 0 rdg 4dgts 100 V 1 mV 10 mV 100 mV 1 V 06 ELECTRICAL SPECIFICATIONS Accuracy is readings digits number of digits or otherwise o o specified at 23 C 5 C less than 75 R H Maximum Crest Factor 1 75 1 at full scale 3 5 1 at half scale with frequency components within the specified frequency ...

Page 6: ...rdg 3dgts 2 0 rdg 5dgts Accuracy Resolution Range 10 00A 1 2 rdg 3dgts Burden Voltage 0 15mV A 0 15mV A 3 3mV mA 3 3mV mA 0 03V A 0 0V A 0 1 A 1 A 10 A 100 A 1 mA 10 mA 07 CAPACITANCE 3 5 6dgts 500 0nF 5 000 F 50 00 F Resolution Range Accuracy 500 0 F 3000 F Additional 50 00nF range accuracy is not specified Accuracies with film capacitor or better Specified with battery voltage above 2 8V approxi...

Page 7: ...4000 counts LCD display 2 Push buttons for special functions features 3 Selector to turn the Power On or Off and select a function 4 Input Jack for 10A 20A for 30sec current function 5 Input Jack for all functions EXCEPT current A mA A function 6 Common Ground reference Input Jack for all functions 7 Input Jack for milli amp and micro amp functions V PRODUCT DESCRIPTION This user s manual uses onl...

Page 8: ...ons Before and after hazardous voltage measurements test the voltage function on a known source such as line voltage to determine proper meter functioning 2 DC 400 0mV range is designed with 1000M high input impedance for least current drain in measuring small signals and can cope better with most commercially available voltage output transducers and adapters The non zero display reading is normal...

Page 9: ...than that indicates a leaky diode defective A zero reading indicates a shorted diode defective An OL indicates an open diode defective Reverse the test leads connections reverse biased across the diode The digital display shows OL if the diode is good Any other readings indicate the diode is resistive or shorted defective Diode Test Capacitance Function Defaults at Press SELECT button momentarily ...

Page 10: ...umption sleep mode auto matically to extend battery life after approximately 30 minutes of no rotary switch or push button operations To wake up the meter from sleep mode press any button momentarily or turn the rotary switch to an adjacent position Always set the rotary switch to the OFF position manually when the meter is not in use Manual or Auto ranging Press the RANGE button momentarily to se...

Page 11: ...easuring functions through this terminal will then be open circuit Refer to the LIMITED WARRANTY section for obtaining warranty or repairing service 18 Battery and Fuse replacement Battery use Standard 1 5V AAA Size NEDA 24A or IEC LR03 battery X 2 Fuses use Fuse FS1 for AmA Current input 0 4A 1000V ac dc IR 30kA F Fuse or better Dimension 6 x 32 mm Fuse FS2 for A Current input 11A 1000V ac dc IR ...

Page 12: ...a KUSAM MECO authorized dealer This warranty does not apply for damaged Ic s fuses burnt PCB s disposable batteries carrying case test leads or to any product which in KUSAM MECO s opinion has been misused altered neglected contaminated or damaged by accident or abnormal conditions of operation or handling KUSAM MECO authorized dealer shall extend this warranty on new and unused products to end us...

Reviews: