MAINTENANCE :

WARNING :

To  avoid  electrical  shock,  disconnect  the  meter  from  any  circuit,
remove  the  test leads from the input jacks and turn OFF the meter
before opening the case. 

Cleaning and Storage :

Periodically wipe the case with a damp cloth and mild detergent; do

not  use  abrasives  or  solvents.  If  the  meter  is  not  to  be  used  for

periods  of  longer  than  15  days,  remove  the  battery  and  store  it

separately. 

Trouble Shooting :

If  the  instrument  fails to operate,  check batteries and test leads, 

etc., and  replace  as  necessary.  Double  check operating 

procedure as described in this manual.

If the instrument voltage-resistance input terminal has subjected to 

high  voltage  transient  (caused  by  lightning  or  switching  surge  to 

the  system)  by  accident  or  abnormal  conditions  of  operation,  the 

series fusible resistors will be blown off (become high impedance) 

like fuses to protect the user and the instrument. Most measuring 

functions through this terminal will then be open circuit.  Refer to 

the  LIMITED  WARRANTY  section  for  obtaining  warranty  or 

repairing service. 

18

Battery and Fuse replacement :

Battery use :

 

     Standard 1.5V AAA Size (NEDA 24A or IEC LR03) battery X 2 

Fuses use :

Fuse (FS1) for

 

AmA

 Current input : 

    0.4A/1000V ac & dc, IR 30kA, F Fuse or better
    

Dimension :

 6 x 32 mm

Fuse (FS2) for A Current input : 
   11A/1000V ac & dc, IR 20kA, F Fuse or better
    

Dimension :

10 x 38 mm

Battery replacement with 

battery access door:

 

Loosen the 2 screws from 

the battery access door of 

the case bottom. Lift the 

battery access door and 

thus the battery compartment

up. Replace  the battery. 

Re-fasten the screws. 

Loosen the 4 screws from the case bottom. Lift the end of the case 

bottom nearest the input jacks until it unsnaps from the case top. 

Replace the blown fuse(s) and/or the battery. 

Fuse replacement (and 
also Battery replacement 
for splash proof version 
without battery access 
door): 

Replace  the  case  bottom,  and  ensure  that  all  the  gaskets  are 

properly seated and the two snaps on the case top (near the LCD 

side) are engaged. Re-fasten the screws. 

Summary of Contents for KM 805s

Page 1: ...h Resistance Tester Digital Panel Meters DC Power Supplies High Voltage Detector Calibrators Gas Analysers Frequency Counter Function Generator Phasing Sticks Battery Tester Waterproof Pen Testers Solar Power Meter EMF Detector Wood Paper Grain Moisture Meter Transistorised Electronic Analog Digital Insulation Resistance Testers upto 10 KV Digital Sound Level Meter Sound Level Calibrator Digital c...

Page 2: ...tal Multimeter TABLE OF CONTENTS TITLE PAGE NO I Safety II Cenelec Directives III Special Features IV General Specifications V Electrical Specifications VI Product Description VII Operation VIII Maintenance IX Test Certificate X Warranty 01 03 03 03 06 10 11 17 19 20 ...

Page 3: ...nstallation Examples are measurements on household appliances portable tools and similar equipment 02 Terms in this manual WARNING Identifies conditions and actions that could result in serious injury or even death to the user CAUTION Identifies conditions and actions that could cause damage or malfunction in the instrument WARNING To reduce the risk of fire or electric shock do not expose this pr...

Page 4: ... Operating below 2000m Pollution degree 2 IV GENERAL SPECIFICATIONS 04 o o Storage Temperature 20 C to 60 C 80 R H With battery removed Temperature Coefficient nominal 0 15 x specified accuracy C o o o o 0 C 18 C or 28 C 40 C or otherwise specified Low Battery Below approx 2 4V Power supply Standard 1 5V AAA Battery x 2 Power Consumption 3 2mA typical Sleep mode Timing Idle for 30 minutes Sleep mo...

Page 5: ... for 400 0 mV range 0 3 rdg 4dgts 0 5 rdg 3dgts 0 5 rdg 3dgts 0 5 rdg 3dgts Accuracy Resolution Range 1000 V 1 0 rdg 4dgts 100 V 1 mV 10 mV 100 mV 1 V 06 ELECTRICAL SPECIFICATIONS Accuracy is readings digits number of digits or otherwise o o specified at 23 C 5 C less than 75 R H Maximum Crest Factor 1 75 1 at full scale 3 5 1 at half scale with frequency components within the specified frequency ...

Page 6: ...rdg 3dgts 2 0 rdg 5dgts Accuracy Resolution Range 10 00A 1 2 rdg 3dgts Burden Voltage 0 15mV A 0 15mV A 3 3mV mA 3 3mV mA 0 03V A 0 0V A 0 1 A 1 A 10 A 100 A 1 mA 10 mA 07 CAPACITANCE 3 5 6dgts 500 0nF 5 000 F 50 00 F Resolution Range Accuracy 500 0 F 3000 F Additional 50 00nF range accuracy is not specified Accuracies with film capacitor or better Specified with battery voltage above 2 8V approxi...

Page 7: ...4000 counts LCD display 2 Push buttons for special functions features 3 Selector to turn the Power On or Off and select a function 4 Input Jack for 10A 20A for 30sec current function 5 Input Jack for all functions EXCEPT current A mA A function 6 Common Ground reference Input Jack for all functions 7 Input Jack for milli amp and micro amp functions V PRODUCT DESCRIPTION This user s manual uses onl...

Page 8: ...ons Before and after hazardous voltage measurements test the voltage function on a known source such as line voltage to determine proper meter functioning 2 DC 400 0mV range is designed with 1000M high input impedance for least current drain in measuring small signals and can cope better with most commercially available voltage output transducers and adapters The non zero display reading is normal...

Page 9: ...than that indicates a leaky diode defective A zero reading indicates a shorted diode defective An OL indicates an open diode defective Reverse the test leads connections reverse biased across the diode The digital display shows OL if the diode is good Any other readings indicate the diode is resistive or shorted defective Diode Test Capacitance Function Defaults at Press SELECT button momentarily ...

Page 10: ...umption sleep mode auto matically to extend battery life after approximately 30 minutes of no rotary switch or push button operations To wake up the meter from sleep mode press any button momentarily or turn the rotary switch to an adjacent position Always set the rotary switch to the OFF position manually when the meter is not in use Manual or Auto ranging Press the RANGE button momentarily to se...

Page 11: ...easuring functions through this terminal will then be open circuit Refer to the LIMITED WARRANTY section for obtaining warranty or repairing service 18 Battery and Fuse replacement Battery use Standard 1 5V AAA Size NEDA 24A or IEC LR03 battery X 2 Fuses use Fuse FS1 for AmA Current input 0 4A 1000V ac dc IR 30kA F Fuse or better Dimension 6 x 32 mm Fuse FS2 for A Current input 11A 1000V ac dc IR ...

Page 12: ...a KUSAM MECO authorized dealer This warranty does not apply for damaged Ic s fuses burnt PCB s disposable batteries carrying case test leads or to any product which in KUSAM MECO s opinion has been misused altered neglected contaminated or damaged by accident or abnormal conditions of operation or handling KUSAM MECO authorized dealer shall extend this warranty on new and unused products to end us...

Reviews: