 To measure tubular object, change the direction

     of the probe for 90°, look for the minimum vaue

=

 For rough surface, don’t press the probe when

     turning it or slipping it, to protect the probe.

=

 When battery voltage is too low to work, a battery

   symbol appears on the screen. It reminds you to
    change new batteries

=

 When you do not use the meter for a long time, 

     please take out batteries to avoid damaging the
     meter by electrolyte.

=

 Don’t store the meter in a high temperature or wet

     place.

=

 After power off, to turn power on again, wait for two

    seconds.

=

 Rough surface may affect test, Burnish it by sand

     paper, and then measure.

=

 When the surface has thick paint, scrape it off 

     and then measure.

=

  Coupling  paste  is  necessary  for  test,  it  can 

     eject air out off the gap between the probe and
    the object  to get ultrasonic  wave  to pass. For
      rough  surface,  more  viscous  coupling  paste 
     must be used, such as lubricating oil.

05

06

“Adjust” 

can  release.  You  can  see  the  bars 

disappear gradually. Finally “4.0mm” displays, the
meter will be calibrated.

7) Notices

=

When  measuring  thickness  under  2.5mm,

    sometimes an error named as “double refraction”
    may  occur.  Another  error  is  named  as  “pulse 
   envelope or cycle jump”. They bring on a bigger
    result than actual thickness. So when measuring
    thin object, should do it repeatedly to look for the
      least  value.  If  necessary,  make  a  thin  block  to
     compare measuring result.

=

The  ultrasonic  wave  velocity  listed  in  Table  (1)  is

    for  reference.  Because  the  component  of  your
    material  may  not  be  identical  with  that  listed  in 
    Table (1), user had better make sample block to test
     the wave velocity in it. 

=

 It is not a good ultrasonic wave conductor to the

    poriferous material, or uncompacted material, it
    can not be measured by the meter. Casting has
    large crystals, if it surface is rough; it is difficult
     for user to test thickness by the meter. 

6) Calibrating the meter

=

When  changing  a  new  probe,  or  the  probe  is 

    abrased  or  the  meter  is  used  for  a  long  time,  you
   need to calibrate the meter by yourself, the method
  is as follows: Power on the meter, smear a little of 
    coupling  paste  on  the  surface  of  the  metal  block
   built -in the shell of the meter. Put the probe on and
  press  it,  when  coupling  symbol  appears,  press
   button 

“Adjust” 

for two seconds, four bars appear

   on the screen. Keep pressing the probe button