07
08
ELECTRICAL SPECIFICATIONS : KM 5040-T
Accuracy is ± (% reading number of digits) or otherwise
specified @ 23°C ± 5°C & less than 75% R.H. TRMS ACV & ACA
accuracies are specified from 5% to 100% of range or otherwise
specified. Maximum Crest Factor <4.2:1 at full scale & <2.1:1 at
half scale, and with frequency spectrums, besides fundamentals,
fall within the meter specified AC bandwidth for non-sinusoidal
waveforms.
AC VOLTAGE
50Hz -- 60Hz
50.00 mV
500.0 mV
5.000 V
50.00 V
500.0 V
1000 V
40Hz -- 500Hz
50.00 mV
500.0 mV
5.000 V
50.00 V
500.0 V
1000 V
±(0.5%rdg + 3dgts)
±(0.8%rdg + 3dgts)
±(1.0%rdg + 4dgts)
±(1.2%rdg + 4dgts)
10
m
V
100
m
V
1 mV
10 mV
100 mV
1 V
Range
Accuracy
Resolution
10
m
V
100
m
V
1 mV
10 mV
100 mV
1 V
SPECIAL FEATURES :
˜
Basic DC Voltage Accuracy :
0.08%
˜
50ms record MAX-MIN readings at fast 20/second
measurement mode
˜
0.8ms Crest (instantaneous Peak-HOLD)
˜
Data Hold & Range- Hold
˜
Backlighted Display
˜
Autoranging
˜
MAX-MIN readings
˜
Fire retarded casing.
˜
Diode & Continuity Test
˜
Type-K Temperature
TM
˜
Audible & Visible BeepJack Input warning against
improper plug in.
˜
Relative-Zero offset mode
˜
Zoom 5x Analog pointer at 60/s
˜
Auto power off
˜
PC Interface (BRUA-85Xa) (Optional)