05
06
GENERAL SPECIFICATIONS
í
Sensing :
AC True RMS
í
Display :
3-4/5 digits 5000 counts LCD display
í
Update Rate :
Fast data : 5 / second nominal;
52 Segment Bar-graph : 60 per second nominal.
í
Operating Temperature :
0°C to 45°C
í
Relative Humidity :
Maximum 80% R.H. for Temperature upto 31°C
decreasing linearly to 50% R.H. at 45°C
í
Storage Temperature :
-20°C to 60°C, 80% R.H. (With battery removed)
í
Pollution degree :
2
í
Altitude :
Operating below 2000m
í
Temperature Coefficient :
Nominal 0.15 x (specified accuracy) / °C
@ (0°C ~ 18°C or 28°C ~ 45°C), or otherwise specified.
í
Power Consumption :
4.3mA typical
í
Low Battery :
Below approx. 7V
í
APO Timing :
Idle for 17 minutes.
í
APO Consumption :
50
m
A typical
í
Power Supply :
Single 9V battery.
í
Dimension :
Approx.186(L)mm x 87(W)mm x 35.5(H)mm;
198(L)mm x 97(W)mm x 55(H)mm with holster
í
Weight :
Approx.340gm, Approx. 430gm with holster.
˜
OPTIONAL ACCESSORIES :
BRUA-85Xa PC interface Kit ( BC - 85Xa RS232 optical adapter cable +
Software CD + BUA-2303 USB - to - Serial adaptor) Bkb32 banana plugs to
K-type socket plug adaptor.
˜
ACCESSORIES :
Test leads (pair), Holster, Battery, User's Manual, Bkp60 banana plug K Type
Thermocouple.
SAFETY
˜
Safety :
Double insulation per IEC / UL / EN61010-1 2nd Ed., 3.0, IEC /
EN61010-2-030 Ed. 1.0, IEC / EN61010-2-033 Ed. 1.0, IEC / UL /
EN61010-031 Ed. 1.1 & CAN / CSA-C22.2 No.61010-1-12 Ed. 3.0 to
CAT III 1000V AC & DC and CAT IV 600V AC & DC.
˜
E. M. C. :
Meets EN61326-1:2013
In an RF field of 3V/m :
Capacitance function is not specified.
Other function ranges : Total Accuracy = Specified Ac 100 digits
Performance above 3V/m is not specified.
˜
Transient Protection :
8kV lightning surge (1.2/50
m
s)
˜
Terminals (to COM) Measurement Category :
V / A / mA
m
A : CAT III 1000 Volts AC & DC, & CAT IV 600 Volts AC & DC.
˜
Overload Protections :
m
A & mA :
0.44A/1000V, IR 10kA , F fuse;
A : 11A/1000V, IR 20kA, F fuse
V : 1100V DC/AC rms
mV,
W
& Others : 1000V DC/AC rms