19
Insufficient ground (must be at least 4 times the detector area)
Flame lifting off burner head (ground), or not continuously in
o
Temperature in excess of 600 F (316 oC) at the flame electrode
●
●
contac t with the flame rod
●
insulato r causing short to ground.
Capacitance & Temperature
Inputs are made via the test lead terminals
COM
/
+
. Defaults at
Capacitance. Press the SELECT button momentarily to select °C
(Celsius) and
°F
(Fahrenheit) in sequence (
°F
selection can be left
out as factory calibration default for countries that only accept metric
units).
20
CAUTION
Discharge capacito r (s) before making capacitance measurements.
Large valu e capacitors should be discharged through an
appropr i a t e resistance load . Using Capacitance function in a live
circuit will produce false resul t s and may damage the meter. In
many cases the suspected comp o nent (s) must be disconnected
from the circuit to obtain accurate measurement readings.
Be sure to insert the banana- plug type -K temperature bead-probe
Bkp60 with correct polarities. Banana-pins to type -K socket
adapter Bkb32 (Optional purchase) can be used to accept other
type-K probes using standard miniature plugs.
Temperature accuracies assume meter interior has the same
temperature (isothermal stage) of the ambient, particularly the plug
of the probe being used, for a correct junction voltage compensation.
Allow the meter interior temperature to atch up with that of the plug
after a sudden change of measuring environmen t and hence the
ambient temperature. This can take up to an hou, r for changes > 5°C,
within a low ventilated sturdy meter housing. The uncompensated
temperature differences, if any, will be reflected as off sets on the
meter readings.
HOLD
HOLD
feature freezes the displa y for later view. LCD turns on.
Press the
HOLD
button momentarily to toggle the feature.
“ ”
Note
Relative
Relative
mode allows the user to offs et the meter consecut ive
measurements wi th the main - display displaying reading as the
reference value . LCD “ ” turns on. Press
(HOLD)
the button
for one second or more to toggle mode.
Relative Mode