06
ELECTRICAL SPECIFICATIONS :
Accuracy is ± (% of reading digits + number of digits) or otherwise
o
o
specified, at 23 C ± 5 C. Maximum Crest Factor <2 :1 at full scale &
<4:1 at half scale or otherwise specified, and with frequency spectrum
not exceeding the specified frequency bandwidth for non-sinusoidal
waveforms.
Range
TM
CLAMP-ON AmpTip AC CURRENT
1)
Accuracy
Resolution
50Hz ~ 60Hz
±(1.0%rdg + 5dgts)
0.01 A
60.00 A
1)
Induced error from adjacent current-carrying conductor :
< 0.01A/A
REGULAR AC CLAMP-ON CURRENT
1)
Induced error from adjacent current-carrying conductor :
< 0.1A/A
Range
Resolution
50Hz ~ 100Hz
0.1 A
600.0 A
100Hz ~ 400Hz
0.1 A
600.0 A
1)
Accuracy
±(1.5%rdg + 5dgts)
±(2.0%rdg + 5dgts)
REGULAR DC CLAMP-ON CURRENT
1)
Induced error from adjacent current-carrying conductor :
< 0.1A/A
2)
Specified with DC-Zero mode applied to offset the non-zero residual
readings, if any.
Range
1) 2)
Accuracy
Resolution
±(2.0%rdg + 5dgts)
0.1 A
600.0 A
07
600.0 V
±(1.0%rdg + 5dgts)
DC VOLTAGE
Accuracy
Resolution
Range
0.1 V
Input Impedance :
10M
, 100pF nominal
AC VOLTAGE (with Digital Low-Pass Filter)
Input Impedance :
10M
, 100pF nominal
Range
Accuracy
Resolution
50Hz ~ 60Hz
±(1.5%rdg + 5dgts)
0.1 V
600.0 V
Range
TM
CLAMP-ON AmpTip DC CURRENT
1)2)3)
Accuracy
Resolution
±(1.0%rdg + 5dgts)
0.01 A
60.00 A
1)
Induced error from adjacent current-carrying conductor :
< 0.01A/A
2)
Specified with DC-Zero mode applied to offset the non-zero residual readings, if any.
3)
Add 5d to the specified accuracy @<4A.
CREST (Peak-Hold)
Applicability
Voltage & Non-invasion Current
functions
Accuracy
Add ± 250 digits to specified accuracy
for changes > 5ms in duration.
3.000 V
±(1.5%rdg + 5dgts)
TM
BeepLit DIODE TESTER
1)
Accuracy
Resolution
Range
1 mV
Test Current :
0.3mA typically
Open Circuit Voltage :
< 3.5VDC typically
Short Beep Alert Threshold :
Drop across 0.850V
TM
BeepLit ON Threshold :
<0.100V
Audible indication :
Beep Sound
Visible Indication :
LCD Backlight.