Chapter 1:  Introduction 

Basic NCS User Manual 

 

12 

 

 

1.4 Wiring Layout 

 

 

 

 

 

LAN cables must comply with the Ethernet specifications(10/100Base-T. 

 

The network wires should not be close to power lines in the hospital. 

 

The power mains should have emergency power backup. 

 

Cavity dimension of 7 x 7 cm is needed for flush installation of Corridor status lights, 

Cancel/Presence, Patient panel and Pull cord switches. 

 

 

Summary of Contents for Basic Nurse Call System

Page 1: ...1 OPERATOR S MANUAL Basic Nurse Call System KTC Medical Co Ltd...

Page 2: ...td No 44 Moayyer St Mofateh Ave TEHRAN IRAN P C 1596785613 Tel 98 21 4336 8000 Fax 98 21 8889 2859 After Sales Services Tel 98 21 4336 8000 Cell 98 912 0854106 Fax 98 21 8889 2859 Web www kt co com Em...

Page 3: ...pt close to it to be used whenever necessary Intended Audience This manual is provided for facility staff members and the patients The users of this manual are expected to have working knowledge of si...

Page 4: ...vides useful information and recommendations about devicefunction Explanation of Symbols use on modules This symbol means that consult user manual of the monitor and pay attention tothe warnings and c...

Page 5: ...n safety of applied parts as there is no floating contact to the patient Operation in Oxygen rich and in the vicinity of flammable anesthetics for KTC Medical Basic Nurse Call System is considered as...

Page 6: ...5 Pear Push Switch 17 06 Pull Cord Nurse Call Switch 18 07 Cancel Presence Switch 19 08 Corridor Status Light 20 09 Care Cleaning PM 21 10 Troubleshooting 28 This guide describes all features and func...

Page 7: ...Chapter 1 Introduction Basic NCS User Manual 7 Chapter 1 Introduction Section Title Page 1 1 General Warnings 8 1 2 Getting started 10 1 3 General Information 11 1 4 Wiring Layout 12...

Page 8: ...by manufacturer The operator must check that system and accessories function safely and see that it is in proper working condition before being used Alarm must be set according to different situations...

Page 9: ...ed adjacent to orstacked with other equipment and that if adjacent or stacked use is necessary the equipment should be observed to verify normal operation in the configuration in which it will be used...

Page 10: ...supply and power cable connection If the problem still exists contact thelocal After Sale Service Power on the Terminal Box Press POWER key on the Terminal box to power on the system Please check the...

Page 11: ...alling the nurses or other staff working in a hospital or other similar facilities considering the priority order defined by time of the call Basic Nurse Call System is a low cost system with 120 ente...

Page 12: ...Ethernet specifications 10 100Base T The network wires should not be close to power lines in the hospital The power mains should have emergency power backup Cavity dimension of 7 x 7 cm is needed for...

Page 13: ...al speaker with melodies stored on a local Micro SD card CUN 02 has a queue management of the first 120 incoming calls in a maximum 240 bed facility It shows the first 4 calls in chronological order W...

Page 14: ...incoming Calls in Chronological order Speaker To play the chime for call alert RJ45 Socket To connect to the Nurse Call System Network Input DC power Socket 5 V DC Power from the external adapter Micr...

Page 15: ...0 2 A 50 HZ Color Grey Port RJ45 connector to Display Corridor CUN 02 Operating Temperature 10 c 50 C Operating Humidity 30 80 RH Mounting Surface Max 3 meters from Staff Console Number Item Descripti...

Page 16: ...ght PBCN 13 61 g PBC 12 68 g PBCN 11 69 g Dimensions W 80mm H 80mm D 20mm Raw Material Polycarbonate ABS Color White Connection 5 Pin Military Gx16 Male Power Supply 12 V Port 5 Pin Male Gxl 6 5 Milit...

Page 17: ...nection 5 Pin Military Gx16 Female Cable Length 4m Operating Temperature 10 C 50 C Operating Humidity 30 80 RH Mounting Tabular at the bedside Wall Number Item Description Nurse Call Switch To initiat...

Page 18: ...mes with a Nurse call switch on the panel to be used as an alternative to the Pull cord one Weight 58 g Dimensions W 80mm H 80mm D 20 mm Raw Material Polycarbonate ABS Color White Power Supply 12 V Co...

Page 19: ...PBC 12 The nurse will use CBN 11 after a call has been taken care of and there is no more activated call associated with the room Weight 58 g Dimensions W 80mm H 80mm D 20mm Raw Material Polycarbonate...

Page 20: ...ght is OFF when no Nurse Call has been activated or after Cancel Presence Switch has been activated after the call Weight 70g Dimensions W 80mm H 80mm D 60mm Raw Material Polycarbonate ABS Color White...

Page 21: ...ries Check if all the power cable and accessories are firmly connected Check all the functions of keyboard and modules to make sure that the monitor is in proper condition If you find any damage on an...

Page 22: ...pay special attention to the following items All the modules in the system shall be kept dust free Do not use strong solvents such as acetone or ammonia Most cleaning agents must be diluted before us...

Page 23: ...veto rough cleaning methods than the housing Don t spray a liquid directly on the screen 9 2 4 Accessories Accessory holders and extension cables should be cleaned and disinfected after each patient o...

Page 24: ...In between patients and as required wipe gently using a moist cloth and warm soapywater or mild detergent In between patients and as required with Alcohol 70 Isopropyl Alcohol N Propanol If needed ca...

Page 25: ...stem and all repairs shall be done by the manufacturer 9 3 1 Storage The storage environment shall be clean and dry If possible use the original packaging of the device If any module falls from falls...

Page 26: ...s 9 3 3 Monthly checks Visual inspection of device Device cleanness Function of keys and indicators Visual inspection of accessories The preventive maintenance PM checklist KTC PM CHK BSC NCS should b...

Page 27: ...ding to the user manual 2 Keyboard Correct function 3 Touch Correct function 4 Display screen Correct display of information 5 Alarms Alarm activation Clarity of alarm sound Correct function of alarm...

Page 28: ...any responsibility for any possible hazard to the patient and the monitor Troubleshooting guide is intended to help users to solve minor problems caused by incorrect use of any module When you face a...

Page 29: ...g Check the cabling Self triggering in a specific room Short in Pear push switch Replace the unit Patient panel is broken Replace the panel Pear push connection Console is broken Check the pear push p...