KMPS-1-64 Manual
12/18
KM 8000 Rev. A
Page 13 of 40
Copyright © 2018 Kulite Semiconductor Products, Inc.
All Rights Reserved.
DYnamic ON/OFf
Sets or returns if dynamic IP acquisition is ON or Off. If on then the
KMPS will first attempt DHCP and if that fails it will use AUTO-IP. If off
it will use the IP address in memory.
Set in programming mode only. Must reset to take effect.
GAteway
xxx.xxx.xxx.xxx
Sets or returns the default gateway
Set in programming mode only.
IP
xxx.xxx.xxx.xxx
Sets or returns the IP address. If set will automatically switch the static
IP.
Set in programming mode only. Must reset to take effect.
IP STream
xxx.xxx.xxx.xxx
Sets or returns the IP address to which the unit will send streaming
data(i.e. the unit's streaming remote IP) Must be set to 0.0.0.0 for TCP
streaming.
Set in programming mode only. Must reset to take effect.
IP REsponse
xxx.xxx.xxx.xxx
Sets the IP address to which the unit will send command responses to
UDP commands (i.e. the unit's response remote IP). If set to 0.0.0.0 then
the scanner will respond to the IP address of the sender.
Set in programming mode only. Must reset to take effect.
MAc
Returns the MAC address
MAc STream
Sets or returns the MAC address to which the unit will send streaming
data. This setting is only used if the IP Stream is set to a fixed value. In
that case the unit will use a static MAC address rather than finding the
MAC address using ARP packets. If the MAC address does not
correspond to the correct IP address the packets may be misdirected by
a network.
MAC address is entered as six pairs of hexadecimal digits separated by
a : or
– (01:23:45:67:89:AB). Set to 00:00:00:00:00:00 for dynamic ARP
(default).
Set in programming mode only. Must reset to take effect.
POrt x
Sets or returns the KMPS local port on which the unit will listen for
commands.
Set in programming mode only. Must reset to take effect.
POrt REsp x
Sets or returns the port on which the unit will respond to commands (i.e.
the unit's response remote port). If set to 0 then the unit will respond to
commands on the same port is was sent from. In UDP broadcast this