5.3 Status messages and diagnostic information
Measurements out of specification
Maintenance
Message
Message
Message
Message
Description
Description
Description
Description
Action
Action
Action
Action
S01
pH value > pH max
pH value out of range
Consider the measuring range limits, otherwise
select a suitable sensor for the process conditions of
the application.
S02
pH value < pH min
pH value out of range
S03
Temp. value > Temp max
Temperature value out of range
S04
Temp. value < Temp min
Temperature value out of range
Message
Message
Message
Message
Description
Description
Description
Description
Action
Action
Action
Action
M01
Slope < -65 mV/pH
Recalibrate the sensor. For more information refer
to
Calibration with PACTwareTM
on page 23.
M02
Slope > -50 mV/pH
M03
Zero point < -99 mV
M04
Zero point > +99 mV
M05
Maintenance interval expired