INSTALLATION
3
15
SMARTPAT PH 8530
www.krohne.com
03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en
Unpacking the sensor
•
Remove by gently twisting and pulling the protective cap from the sensor
1
.
•
Lay the sensor on a soft ESD mat or soft paper tissue
2
.
•
Leave the protection cap on the VP connector, as long as the sensor is not connected to the
cable.
3.4 Installation procedure
Because a new pH sensor needs to be calibrated before it is installed into its final measuring
location, it is important to follow the installation order:
1
Connect the sensor to the junction box or directly to the process control system.
2
Calibrate the sensor.
3
Install the sensor into its final measuring location.
The required steps are explained in the following sections.
Figure 3-1: Unpacking the sensor
WARNING!
During installation of the device make sure that you use ESD (electrostatic discharge) protection
equipment.