TECHNICAL DATA
6
27
OPTISYS TSS X050 SERIES
www.krohne.com
07/2018 - 4004781502 - MA OPTISYS TSSx050 R02 en
Measuring accuracy
Accuracy
<
±
5% of the maximum value range
Repeatability
<
±
0.5% of the maximum value range
Response time
1.5 seconds
Operating conditions
Process temperature
0...+90
°
C / +32...+194
°
F
Ambient temperature
-20...+70
°
C / -4...+158
°
F
Max. admissible temperature
+90
°
C / +194
°
F
Max. admissible sterilisation
temperature
+140
°
C / +284
°
F, max. 2 hours
Max. pressure
10 bar / 145 psi at 60
°
C / 140
°
F
Ingress protection
IP69K
Installation conditions
Installation position
Installed on tanks or in pipelines.
Device can be operated in any position but it is recommended with a product flow
from bottom to top.
Optical window of the device is submerged in the process medium
For flushing or removing device during running process, for types TSS 2050/4050
retractable assemblies are optionally available.
Process connection
G1/2 (1050/3050) or PG 13.5 (2050/4050)
Weight
TSS 1050/3050: 700 g / 1.5 lb
TSS 2050/4050: 850 g / 1.9 lb
Materials
Sensor body
Stainless steel 1.4435 (316L)
Surface quality: electro-polished < Ra 0.37
μ
m
Measuring window
Sapphire
Electrical connection
Power supply
12...30 VDC
Load
≤
(U
b
- 4 V)/20 mA (max. 400
Ω
at 12 V, 1000
Ω
at 24 V, 1300
Ω
at 30 V)
Analogue output
4...20 mA
Current demand
Approx. 80 mA (30 VDC and analogue output = 22.5 mA)
Power input
2.4 W max.
Current limit
3.5 mA min.; 22.5 mA max.; adjustable
Switching output
All devices: NO or NC adjustable; PNP- switching; 200 mA max., thermally
protected
Teach input (only for
TSS 1050/2050
Digital input; +12...30 VDC max. 1.6 mA input current
Cable connection
5-pin M12 connector
Cable length
Standard: 5 m / 16.4 ft
Option: 10 m / 32.8 ft, 20 m / 65.6 ft or 30 m / 98 ft