6
TECHNICAL DATA
26
OPTISYS TSS X050 SERIES
www.krohne.com
07/2018 - 4004781502 - MA OPTISYS TSSx050 R02 en
6.2 Technical data
INFORMATION!
•
The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local sales office.
•
Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Downloadcenter).
Measuring system
Measuring principle
180
°
transmitted light absorption method pulsed NIR 880 nm. Absorption of light on
suspended solids and colloidal particles. The light passes the measured particles
between one emitter and one detector.
Application
Total suspended solids, turbidity or optical density (extinction) in fluids in the near
infrared range (880 nm wavelength)
Design
Variants
4 different types
TSS 1050
TSS 1050
TSS 1050
TSS 1050
Simply designed, competitive NIR system with a scale between 0 to 100%
absorbtion characteristics of fluids. G1/2 process connection.
TSS 2050
TSS 2050
TSS 2050
TSS 2050
Simply designed, competitive NIR system with a scale between 0 to 100%
absorbtion characteristics of fluids.
This system is used in a manually (RET) or automatically (RAM) operated
retractable assembly of the SENSOFIT family. PG13.5 process connection.
TSS 3050
TSS 3050
TSS 3050
TSS 3050
Measuring system for monitoring the optical density (extinction) or absorption of
fluids in order to monitor continuous process results or to securely indicate
changes. This device has the possibility to be calibrated to improve accuracy. G1/2
process connection.
TSS 4050
TSS 4050
TSS 4050
TSS 4050
Measuring system for monitoring the optical density (extinction) or absorption of
fluids in order to monitor continuous process results or to securely indicate
changes. This device has the possibility to be calibrated to improve accuracy.
This system is used in a manually (RET) or automatically (RAM) operated
retractable assembly of the SENSOFIT family. PG13.5 process connection.
Display (option)
Device can be parameterised using the function keys on the display.
Light method specifications
Wavelength: 880 nm
Light source: LED
Measuring range
Measuring range
Measuring range
Measuring range
TSS 1050/2050
Scale: 0...100%
OPL 5 mm / 0.2": 0...500 EBC / 0
…
2000 FAU / 5.0 g/l ~ 0.4% TSS*
OPL 10 mm / 0.4": 0...250 EBC / 0
…
1000 FAU / 2.5 g/l ~ 0.2% TSS*
OPL 20 mm / 0.8": 0...100 EBC / 0
…
400 FAU / 1.0 g/l ~ 0.1% TSS* (only TSS 1050)
* the values represent about 80% of the display scale
TSS 3050/4050
0...3 AU (Absorption Units)
0
…
6 OD (Optical Density)
0
…
13000 FAU (Formazine Absorption Unit)
0
…
13000 FTU (Formazine Turbidity Unit)
0...3250 EBC (European Brewery Convention)
0
…
26.65 g/l (grams per litre)