Manual OPTISENS CAC 050 / IAC 050
OPTISENS CAC 050 / IAC 050
14
4.
Adjustments for the measurement
Main menu
Temp. comp.
.
Selection between automatic and manual temperature compensation,
adjustment of temperature for manual compensation.
.
.
Basic settings
Basic settings
Cal. temp.
Calibration of the temperature measurement.
.
.
Temp. coeff.
Adjustment of the deviation per °C
Cell factor c
Adjustment of the cell factor (sensor charact.)
Cable comp.
Zero point correction (cable capacitance)
.
Meas. range
Selection of the measuring range
Averaging
Activation of a smoothing function
The conductivity measurement is influenced by temperature. This influence is compensated automatically or
manually. For manual compensation the temperature is entered manually, for automatic compensation a
temperature sensor must be connected. Compensation is carried out linearly using a temperature coefficient
in %/K which depends upon the composition of the test water.
The capacitance of the sensor cable can lead to a slight deviation of the measurement. This deviation can be
eliminated by cable compensation.
Especially in the lower measuring ranges the measurement tends to get a bit unsteady due to the high
resistance of the water. The signal can be smoothed out by activating an averaging function.
Conductivity sensors are optimized for certain measuring ranges by their geometric dimensions. These
define the cell factor (c value). If you want to use a sensor with a different cell factor or if you want to
calibrate the measurement, adjust the cell factor accordingly.
All these adjustments are part of the basic settings, since they are carried out only once at the beginning.
4.1
Configuration of the converter
Main menu
.
.
Basic settings
Basic settings
.
.
Configuration
Configuration
Measurement
Temp. sensor
The devices CAC 050 and IAC 050 2 analog inputs:
Input 1 is for conductivity measurements (conductive with CAC 050, inductive with IAC 050) for a variety of
measuring ranges:
Method
Measuring range
Sensor
conductive
0.00...20.00 MOhm/cm
c = 0.05/cm with Pt100
0.000...2.000 µS/cm
c = 0.05/cm with Pt100
0.00...20.00 µS/cm
c = 0.05/cm with Pt100
0.0...200.0 µS/cm
c = 0.05/cm with Pt100
0.000…2.000 mS/cm
c = 0.2/cm with Pt100
0.00...20.00 mS/cm
c = 1/cm with Pt100
0.0...200.0 mS/cm
c = 10/cm with Pt100
0.00...20.00 mS/cm
Sensor with NTC
0.0...200.0 mS/cm
Sensor with NTC
0...2000 mS/cm
Sensor with NTC