Manual OPTISENS CAC 050 / IAC 050
27
OPTISENS CAC 050 / IAC 050
9.
Technical data
9.1
Technical data
Feature
CAC 050 R
IAC 050 W
CAC 050 W
IAC 050 W
Version
panel-type converter
wall-mounting enclosure
Dimensions (w x h x d)
96 x 96 x 127 mm / 3.8 x 3.8 x 5”
165 x 160 x 80 mm / 6.5 x 6.3 x 3.1”
Weight
0.8 kg / 1.8 lbs
1.0 kg / 2.2 lbs
Terminals
screw terminals max. 1.5 mm
2
Spring terminals max. 1,5 mm
2
Protection class
IP 54 (front), IP 55 (front door)
IP 65
Power supply
230 V +6/-10 %, 50/60 Hz, optional 110 V or 24 V
Internal fuse
none
230 V: 63 mA slow
110 V: 125 mA slow
24 V: 800 mA semi-slow
Power consumption
10 VA
Display
LCD, 2-line, 2x16 characters, illuminated background
measured value and temperature with dimensions indication of relay
status
Current output
0/4...20 mA, galvanically isolated, max. loading 500 Ohm
Interface (option)
RS485, Baud rate 9600, data format 8Bit, 1 Start and 1 Stop bit
Controller
ON/OFF controller with hysteresis, P or PI controller as Pulse- Pause- or
Impulse-frequency controller, steady controller, bidirectional PI control
action, adjustable onset-delay, dosage control function, manual operation
of the relays, controller stop via external switch or level sensor (lack-of-
water sensor)
Set points
2 set points adjustable within the measuring range
Alarm function
with min. and max. limit and turn-on delay
Contact rating
6 A/ 250 V, max. 550 VA resistive load (with RC protective circuit)
Operation temperature
0…+50°C / 32…+120°F
Storage temperature
-20…+65°C / -5…+150°F
Humidity
max. 90% at 40°C /105°F non-condensing