Manual OPTISENS PAC 050
OPTISENS PAC 050 26
8.5
Maintenance of the pH or ORP measurement
The metallic surface of the ORP electrode must be cleaned regularly. pH and reference electrodes change
with time, so their characteristics have to be determined regularly by means of calibration.
NOTE
The instrument checks the sensor’s characteristic parameters after each calibration and displays an error
message if the sensor has to be replaced. See chapter „Error messages“.
8.6
Disposal
For disposal please notice that the device contains electrolyte capacitors which have to be disposed
separately.
8.7
Service
Main menu
.
Unit
no.
.
No.
041
.
Software
date
.
M/Y
1.00
.
Service
Product.
date
Service
Product info
M/Y 1.00
Analog
inputs
Analog inputs
Input 1 003
Analog
inputs
Input 2 25.0°C
Erase
settings
Erase settings
Press
&
In this menu you will find information which is especially important for any inquiries, updates or problems.
8.7.1
Product info
These figures allow a precise identification of the device (hardware and software).
8.7.2
Analog inputs
Here you can see the raw data the device obtains from the sensors. They are not influenced by
compensations or calibration and offer valuable information in case of problems with the measurement or the
device.
If you have problems interpreting this data, send them to your supplier together with the device data and ask
for support.
8.7.3
Erase settings (reset)
With this function you can erase all customer settings and restore the original at-works data.
The process takes some 30 seconds. When it is finished the display will show the measured value, and the
controller will switch off.