Manual OPTISENS PAC 050
21
OPTISENS PAC 050
5.7
Manual operation of the relays
6.99 pH 25.0°C
F
S1
F
S2 Auto
6.99 pH 25.0°C
F
S1
F
S2 Man
1) If the controller is ON, switch it OFF with key
.
Instead of „Auto“ the display shows „Man“.
6.99 pH 25.0°C
S1
F
S2 Man
2) Switch to the operation mode of S1 with key
.
The square to the left of S1 starts to flash.
6.99 pH 25.0°C
J
S1
F
S2 Man
3) Switch ON relay 1 with key
.
The square to the left of S1 gets dark.
6.99 pH 25.0°C
S1
F
S2 Man
4) Switch OFF relay 1 again with key
.
The square gets light.
6.99 pH 25.0°C
F
S1
S2 Man
5) Switch to the operation mode of S2 with key
.
The square to the left of S2 starts to flash.
6.99 pH 25.0°C
F
S1
J
S2 Man
6) Switch ON relay 2 with key
.
The square to the left of S2 gets dark.
6.99 pH 25.0°C
F
S1
S2 Man
7) Switch OFF relay 2 again with key
.
The square gets light.
6.99 pH 25.0°C
F
S1
F
S2 Man
8) Leave the operation mode of relay 2 with key
.
Both squares appear light, none flashes - You have left the operation
mode.
For manual operation you need no menu.
•
With key
you switch OFF the controller.
•
With key
you switch between Manual operation <> operation mode S1 <> operation mode S2 <>
manual operation.
•
In the operation mode you can switch ON and OFF the selected relay with key
.
•
A flashing square indicates that the relay is in operation mode.
•
A dark square indicates that the relay is switched ON.
•
A light square indicates that the relay is switched OFF.
WARNING
If you switch ON a relay it stays ON until you switch it OFF again manually!