13
Fct. Text
Description
and settings
3.0
INSTALL
Main menu 3.0 Installation
3.1.0
BASIS.PARAM.
Submenu 3.1.0 Basisparameters
3.1.1
FULL SCALE
Full-scale range for flowrate Q
100%
Unit:
select from list
under Fct. 3.2.1
Value: 9.5*10
-7
- 150.8 m
3
/Sec or
3.9 - 1,987,200 US Gal/min
After selecting unit, call numerical
value with
↵
key, 1
st
digit flashes.
3.1.2
REV. SCALE
Different range for reverse flow
required? Setting NO or YES
3.1.3 REV.
VALUE Full-scale range for reverse flow
(appears only if
YES
set
under Fct. 3.2.1)
Unit:
select from list
under Fct. 3.2.1
Value: 9.5*10
-7
- 150.8 m
3
/Sec or
3.9 - 1,987,200 US Gal/min
Value must not be larger
than that of Fct. 3.1.1!
After selecting unit, call numerical
value with
↵
key, 1
st
digit flashes.
3.1.4
ZERO SET
Zero setting
•
FIXED.VALUE
•
VALUE.MEASU.
(Carry out only at “zero” flow and
with completely filled measuring
tube).
1) Inquiry:
CALIB. NO
or
YES
2) if YES: calibration (duration
approx. 20 Sec) with zero display
in
PERCENT
of Q
100%
3) Inquiry:
STORE NO
or
YES
.
3.1.5
METER SIZE
Meter size
Unit:
mm
or
inches
Value:
25
-
1200 mm
or
0.98
-
48 inches
After selecting unit, call numerical
value with
↵
key, 1
st
digit flashes.
3.1.6
GK VALUE
Primary head constant GK
(see also primary head nameplate).
Range:
0.5
-
14
3.1.7
FLOW DIR
Define direction of forward flow,
see Sect. 5.4.
Setting:
+
or
−
, acc. to direction of
arrow
on primary head.
3.1.8
MIN SP. S.
Minimum sound velocity
Minimum value used for I
0%
or F
0%
(when function
SOUND.VELO.
selected in 3.3.1 or 3.4.1)
Value: 0 to 5000 m/s
3.1.9
MAX SP. S
Maximum sound velocity
Maximum value used for I
100%
or F
100%
(when function SOUND.VELO.
selected in 3.3.1 or 3.4.1)
Value: 1 - 5000 m/s