Installation and operating instructions
32 BATCHFLUX
5.8 Status output S (option)
No HARDWARE!
Function must be set to “OFF”.
Fct. 1.7 IND. OUTP. S
Press key
→
.
Select function for the status output,
press key
→
.
•
OFF
(switched
off,
no function)
•
ON
(indicates that
flowmeter is operative)
•
V/R INDIC.
(indicates direction for
current output and
pulse output, F/R mode)
•
TRIP.POINT
(setting range: 002 -
115 PERCENT of Q
100%
,
full-scale
range)
endwert)
•
ALL ERROR
(indicates all errors)
•
FATAL.ERROR
(indicates
Fatal Error only)
Transfer to number setting with key
↵
, 1st
digit (cursor) flashes. Change flashing digit
(cursor) with keys
↑
and
↓
, shift cursor 1
place to right or left with keys
→
and
←
.
Press key
↵
to return to
Fct. 1.7 IND. OUTP. S.
Characteristics of
Switch open
Switch closed
status output
OFF
no function
ON
(e.g. operation indicator)
Power OFF
Power ON
F/R INDIC.
F
orward flow
R
everse flow
TRIP.POINT
Inactive Active
ALL ERROR
Error(s) No
error(s)
FATAL. ERROR
Error No
error
For factory settings, please refer to the report on settings and to Sect. 3.2.
Connection diagrams: see Sect. 2.2.
5.9 Language
Fct. 3.1 LANGUAGE
Press key
→
.
Select language for texts in display
•
D
(German)
•
GB
(English)
•
F
(French)
• others available on request
Select with keys
↑
and
↓
.
Press key
↵
to return to Fct. 3.1 LANGUAGE.
5