39
12
1x
short
Quitting the SETUP – Menu
13
1x long
For confirming and saving.
The instrument returns to the last program used
The instrument must be adjusted until the nominal value is reached!
If „
ERR 07
“ appears in the display the OFFSET value is out of the acceptable
range (
Technical Data). The instrument must be returned to the
manufacturer for checking.
You can select between „mm“ and „inch“.
The gage must be switched on. The SETUP – Menu can be activated in each program.
1x long
Activating
SETUP – Menu
3x
short
„
UNIT
“ will appear in the display
1x long
Activating UNIT – Menu
The last used measuring unit will appear in the display /
Factory setting: mm
short
Changing the measuring unit
1x long
For confirming and saving
Afterwards you will return to the SETUP - Menu.
3x
short
Quitting the SETUP – Menu
1x long
For confirming and saving.
The instrument returns to the last program used
After changing the unit the tolerance limit must be checked again, as the tolerance limit will
be rounded automatically to the new unit and could be incorrect.
Zero point adjustment (OFFSET)
Changing the unit of measurement mm / inch (UNIT)