Operating Manual Krantz
VIRUS
PROTECT
Page
22
Subject to technical changes.
Close the front cover by
gently
tightening the 6 hexagon socket captive screws M6. To
do this, lift the front panel with one hand and align it.
Switch on the device, as outlined in section 8.3. After a correct filter change, the
message CHANGE PRE-FILTER ELEMENT should no longer be displayed.
9.2
Changing the HEPA Filter Element
The device’s display indicates when the filter element of the pre-filter stage must be replaced. See
8.6.3.
Replacements for these pre-filter and HEPA filter elements should be procured from Krantz
GmbH, since only these filter elements have an inactivating, antiviral lining! The use of
other filter elements will void any warranty and Krantz cannot accept any liability for any
damage that results! A list of the required spare parts can be found in section 10
“Consumables” or contact us directly:
Krantz GmbH I Uersfeld 24 I 52072 Aachen
Phone: +49 (0) 0241 441- 0