background image

Notes

14

OII 08en-GB Edition 2021-07

Operating instructions

Summary of Contents for EET 32

Page 1: ...Operating instructions KRAL Temperature Sensors EET 32 EET 33 EET 34 EET 38 OII 08en GB Edition 2021 07 Original instructions www kral at ...

Page 2: ...iption 8 4 1 Structure 8 4 2 Functional principle 8 5 Transportation storage 8 5 1 Unpacking and checking the state of delivery 8 5 2 Storing the sensor 8 6 Installation removal and connection 9 6 1 Dangers during installation removal and connection 9 6 2 Installing the EET in OME 9 6 3 Installing the EET in OMG 10 6 4 Installing the EET in OMP 10 6 5 Installing the EET in the piping 11 6 6 Connec...

Page 3: ...ew Target group Activity Qualification Fitter Mounting connection Qualified personnel for mounting Electrician Electrical connection Qualified personnel for electric installation Trained personnel Delegated task Personnel trained by the operator owner who know the task delegated to them and the possible dangers arising through improper behaviour Tab 1 Target groups 1 3 Associated documents o Decla...

Page 4: ...ns The following safety instructions must be observed strictly o Read and observe these operating instructions carefully o Observe the operating instructions for the flowmeter and the electronics o Have work only carried out by qualified trained personnel o Wear personal protective equipment and work carefully o Depending on the operating conditions the service life of the sensors is limited by vi...

Page 5: ... max C 50 260 Tab 2 Operating limits In addition the operating limits of the corresponding flowmeter and the additional sensors must be ob served 3 2 Selecting the temperature sensor Flowmeter Installation in pipeline Size Series OME OMG OMP 13 EET 32 EET 34 EET 38 Any temperature sensor possible 20 EET 32 EET 33 EET 33 32 EET 32 EET 34 EET 38 EET 34 EET 38 52 EET 33 EET 34 EET 38 EET 33 68 EET 34...

Page 6: ... o Housing material 1 4571 o Seal FPM o Connection type Cable with wire ends o Cable sheath Teflon PTFE o Wire cross section mm2 3 x 0 34 o Cable diameter mm 3 3 o Cable length m 3 o Weight g 110 o Degree of protection IP 65 o Tightening torque Nm 30 Dimensions Unit EET 32 EET 33 EET 34 o D mm 4 6 6 o L1 mm 20 25 35 o L2 mm 29 34 44 o B SW 22 The temperature sensor type version is indicated on the...

Page 7: ...ith wire ends o Cable sheath Teflon PTFE o Wire cross section mm2 3 x 0 34 o Cable diameter mm 3 3 o Cable length m 3 o Weight g 110 o Degree of protection IP 65 o Tightening torque Nm 30 General information Use in potentially explosive atmospheres According to 2014 34 EU ATEX o Ex marking ll 2G Ex ia IlC T1 T6 Gb Dimensions Unit EET 38 o D mm 6 o L1 mm 35 o L2 mm 44 o B SW 22 The temperature sens...

Page 8: ...so be used without falsifying the measurement result The temperature sensor is screwed into the hole provided in the flow meter or into the piping 5 Transportation storage 5 1 Unpacking and checking the state of delivery Personnel qualification o Trained personnel 1 Upon delivery check the product for damage during transportation 2 Report damage during transportation immediately to the manufacture...

Page 9: ...ds o Torque wrench WARNING Risk of injury through escaping medium Media can be hot poisonous combustible and caustic Before installing the sensor switch off the system and depressurise it Wear personal protective equipment at all times during operation Pay attention to face protection Collect any escaping medium safely and dispose of it in an environmentally compatible manner in accordance with th...

Page 10: ...d remove the screw plug from the sensor bore 2 Clean the sensor bore No foreign bodies should penetrate the flowmeter 2 or piping system 3 Select suitable sensor 1 Ä Technical data Page 5 4 Screw the sensor into the sensor bore as far as it will go maximum tightening torque 30 Nm 5 Connect the sensor cable observing the connection diagram Ä Technical data Page 5 6 4 Installing the EET in OMP Perso...

Page 11: ...ING Risk of injury through escaping medium Media can be hot poisonous combustible and caustic Before installing the sensor switch off the system and depressurise it Wear personal protective equipment at all times during operation Pay attention to face protection Collect any escaping medium safely and dispose of it in an environmentally compatible manner in accordance with the applicable local regu...

Page 12: ...ct the extension cable observing the connection diagram Ä Technical data Page 5 6 7 Expanding the EET Personnel qualification o Electrician WARNING Risk of injury through escaping medium Media can be hot poisonous combustible and caustic Before installing the sensor switch off the system and depressurise it Wear personal protective equipment at all times during operation Pay attention to face prot...

Page 13: ...nding flowmeter operating instruc tions 1 2 Contacts corroded Check and clean contacts 2 External interference Follow instructions for routing the cables Ä Installation removal and connec tion Page 9 Tab 5 Fault table 9 Accessories 9 1 Junction box The junction box facilitates the electrical connection of the various sensors Up to three sensors can be connected The sensor cables are combined into ...

Page 14: ...Notes 14 OII 08en GB Edition 2021 07 Operating instructions ...

Page 15: ...Notes Operating instructions OII 08en GB Edition 2021 07 15 ...

Page 16: ...KRAL GmbH 6890 Lustenau Austria Tel 43 5577 86644 0 E Mail kral kral at www kral at ...

Reviews: