background image

Operating Instructions

OIE 12en  Edition 2019-02 

7

Temperature compensation

F

u

n

c

ti

o

n

 d

e

s

c

ri

p

ti

o

n

Temperature compensation

If the KRAL volumeter is additionally equipped with a temperature sensor, the current density of the 

flowing fluid can be calculated from this measured value by means of a stored density table.

At the option "Volume at X°", a normalized volume measurement is then possible at which the 
displayed values are converted to a reference temperature X° that can be selected freely. This 
ensures that measuring errors caused by changes in the density due to temperature variations are 

avoided.

The option "Volume at TempA" calculates the volume back to the temperature volumeter A. This 
option allows the comparison with a reservoir level.

Measuring errors are also reduced at the "Mass calculation" option, since the device can now 
process the actual density and not only a stored mean value at the mass conversion. For the case of 
operation with different fluids two different density tables can be entered and selected.

Differential measurement

The KRAL electronic unit can process the signals of two volumeters and determine and display the links 
possible with it.  

The option differential measurement "A-B" allows the subtraction of the values of the two channels, 

e.g. feed and return of a consumer supplied with a closed circular pipeline and thus allows the direct 
display of the consumer of this device.

The option sum "A+B" allows the addition of the two measured values and thus, for example, the 

display of the total of two devices.

Circulation ratio and threshold value

The ratio A/(A-B) is called the "Circulation ratio". On the basis of the laws of error propagation, a strongly 

rising error of the displayed differential value A-B results for the option "Differential measurement" at 
values A/(A-B) >> 1, so that the value A/(A-B) can also be used to judge the reliability of a differential 
measurement. For the case A/(A-B) >> 1, for example in the case of a deactivated consumer but with a 

circulation pump that continues to be operated, a threshold value can be specified for A-B below which 
the measured values are not taken into consideration for the sum calculation.

Averaging

A strongly fluctuating flow rate causes a jumping display or as a result a fluctuating analog output. The 
averaging function reduces this effect by generating a averaging across several measured values. The 

number of measured values for averaging can be set, see "3.06 Averaging analog output", Page 41, and 
see "2.16 Averaging display rate", Page 39.

Limit value

The "Limit value bypass" function allows the automatic activation of a bypass valve when a volumeter 
blocks. The bypass valve is actuated via Relay output 1, see "3.13 Function relay 1", Page 42.

Collective error message

When an error message occurs, Relay output 2 is switched for the collective error.

Summary of Contents for BEM 500 Series

Page 1: ...www kral at Operating Instructions KRAL Electronics Series BEM 500 OIE 12en Edition 2019 02 HW 3 003 SW 3 004...

Page 2: ...electronic unit from the wall 15 Removing the electronic unit from the pipe volumeter 15 Technical data Termination panel of the electronic unit 16 Technical data of connections 17 Power supply 17 Te...

Page 3: ...unction analog output 40 3 02 Allocation analog output 1 40 3 03 Scale analog output 1 40 3 04 Allocation analog output 2 40 3 05 Scale analog output 2 40 3 06 Averaging analog output 41 3 07 Function...

Page 4: ...employees read and observe these instructions and the associated documents in particular the safety instructions and warnings Observe additional system specific directives and regulations Specialist p...

Page 5: ...ons for the Volumeter Corresponding operating instructions for the sensor Calibration certificate Work sheet The following safety instructions must be observed No liability is accepted for damage aris...

Page 6: ...the total flow volume as a total value since the device was last reset Rate values represent the current instantaneous values total values correspond to the sums since the last resetting Mass calculat...

Page 7: ...d with a closed circular pipeline and thus allows the direct display of the consumer of this device The option sum A B allows the addition of the two measured values and thus for example the display o...

Page 8: ...the filling process see 3 13 Function relay 1 Page 42 The filling process can also be interrupted or aborted Electronic evaluation The electronic unit receives signals from the sensors and calculates...

Page 9: ...put 1 Pulse output Fig 2 Basic Temperature compensation 1 Volumeter 1 Pick up 1 Temperature sensor Pt 100 1 BEM 500 electronic unit Electronic evaluation Volume measurement Mass flow measurement Tempe...

Page 10: ...0 electronic unit Electronic evaluation Differential measurement Mass flow measurement Temperature compensation 2 Relay outputs 2 Analog outputs 2 Pulse outputs Fig 7 Basic Flow direction detection 2...

Page 11: ...conditions Page 21 Storage If possible store the electronic unit in original packaging in a cool and dry place while observing the ambient conditions see Ambient conditions Page 21 Disposal Safety in...

Page 12: ...following safety instructions must be observed All installation and removal work may only be carried out by qualified personnel Read the operating instructions and observe the relevant instructions Th...

Page 13: ...g is also possible without a housing cover 6 and seal with cable entries 5 see Fig 2 page 13 An adapter set is available for converting BEM 4U to the electronic unit BEM 500 see Adapter set for conver...

Page 14: ...ll mounting a universal mount is available as an accessory see Accessories mounting page 54 Prerequisite Universal mount is mounted to the wall All cables have been cut to length and connected Aids KR...

Page 15: ...crew the connecting screws between the front and sealing frames 3 Pull of the sealing frame to the rear 4 Pull the electronic unit with front frame from the front out of the control cabinet 5 Press th...

Page 16: ...1 4 20 mA 5 Analog output 2 4 20 mA 6 Pick up A1 7 Pick up B1 8 Analog output 1 0 10 V 9 Analog output 2 0 10 V 10 Pulse output 1 24 V 11 Pulse output 2 24 V 12 Temperature input A Pt100 13 Temperatu...

Page 17: ...uts Minimum limit frequency 0 3 Hz Maximum limit frequency 20 kHz Power supply for pick up 24 V DC or Namur 8 2 V Input impedance Namur 1 k 24 V DC 3 2 k Switching threshold hysteresis Namur 1 65 V 0...

Page 18: ...djustable from 2 200 ms Max output frequency at pulse width 2 ms Mode Independent 250 Hz Mode Encoder 125 Hz Relay outputs Potential free change over contact Nominal load voltage 250 V AC 30 V DC Swit...

Page 19: ...04 Volumeter A total TA1 4100 2 1 3 2000000000 Unit total 0103 4100 0002 D037 Volumeter A total TA2 4102 2 1 3 2000000000 Unit total 0103 4102 0002 71F7 1 05 Volumeter B rate QB 400C 2 1 3 2147483647...

Page 20: ...BEM with the Modbus Modbus function 10 can be used 10 hex write holding register preset multiple register Data exchange via the bus connection is not password protected Avoid unintended overwriting of...

Page 21: ...ndensing EMC emitted interference and immunity to interference EN 61326 Vibration EN 60068 2 47 EN 60068 2 64 Shock EN 61373 Isolation 500 V Degree of protection IP 65 Connection PNP NPN Namur Termina...

Page 22: ...e Serial interface Connection Function Terminal Pulse output 1 Signal 1 Gnd 2 Pulse output 2 Signal 34 Gnd 35 Connection Function Terminal Relay output 1 Bypass valve or filling valve NO contact NO 9...

Page 23: ...elines at the workplace Knowledge of the electrotechnical safety guidelines The following safety instructions must be observed during the electrical installation The following qualifications are requi...

Page 24: ...ar device cover and remove the terminal tool 2 Use the diagonal cutter to adapt the cable entry to the cable diameter 3 Pull the cable through the cable entry 4 Hook the short arbor of the terminal to...

Page 25: ...he side of the electronic unit see Tab 1 page 25 and see Notes in the supplied wiring diagram 8 Connect the cables for temperature sensor s in accordance with the wiring diagram on the electronic unit...

Page 26: ...y connected Aids KRAL tool set Diagonal cutter 1 Remove the rear device cover 2 Adapt the cable entry to the cable diameter and cut it to length see Connecting cables to the tension spring terminals p...

Page 27: ...system is de energized 2 Remove the rear device cover 3 Check that the wiring of the power supply at the termination panel is firm 4 Check the connection of the power supply to the system 5 Check the...

Page 28: ...UP Change to the following page of the menu Select the previous unit Increase the digit ARROW DOWN Change to the previous page of the menu Select the next unit Decrease the digit ARROW RIGHT Navigate...

Page 29: ...eat Steps 1 3 to activate the password Change the password see 2 02 Change password Page 37 1 Enter the password see above 2 Use ARROW UP to change to the page 2 02 Setting Change Password 3 Press SET...

Page 30: ...elected After you have navigated to a different page the language is used in the display Calling up help see 1 31 Help operation Page 34 1 Press ARROW RIGHT and ARROW LEFT simultaneously Information a...

Page 31: ...nsumption since last reset without password protection T2 Total consumption since last reset with password protection QA Current flow Volumeter A supply line QB Current flow Volumeter B return line Te...

Page 32: ...Select unit total 2 07 Select unit temperature 2 08 Select unit density 2 09 Select density determination 2 10 Number of decimal places 2 11 Display start message 2 12 Function pick up 2 13 Function...

Page 33: ...tive values here 3 14 Limit value bypass 3 15 Time delay bypass 3 16 Waiting period repeat bypass 3 17 Relay 1 switch 3 18 Relay 2 switch 3 19 Address Modbus 4 K factors Volumeter A 4 01 Volumeter A P...

Page 34: ...for operating Press ARROW UPor ARROW DOWNto navigate in the text Use SET to exit the help Menu 8 Alarms Shortcut Press SET and ARROW LEFT simultaneously Use ARROW UPor ARROW DOWNto call up the existi...

Page 35: ...since the last resetting Resetting of the value functions analog to menu item 1 02 1 05 Volumeter B Description analogous to menu item 1 03 1 06 Volumeter B Total with reset Description analogous to m...

Page 36: ...age can be called up via the shortcut see 1 13 Setting select language page 34 Menu 2 General settings Menu 2 allows general adjustments in accordance with the requirements of the measuring task All t...

Page 37: ...e unit of the temperature can be selected in accordance with the requirements 2 08 Select unit density The unit of the density can be selected in accordance with the requirements Setting Meaning Volum...

Page 38: ...for push pull pick ups 2 13 Function pulse inputs If one pick up each is used per volumeter the Counter function is selected Encoder is used if the volumeter is equipped with 2 pick ups flow directio...

Page 39: ...e Modbus 2 17 Deactivate alarm messages The display of alarm messages can be suppressed on this page Note The collective error message is then also deactivated The key combination SET and ARROW LEFT C...

Page 40: ...flow Volumeter A TB1 Total flow Volumeter B 3 03 Scale analog output 1 The scale of the analog output is used to set the maximum value The maximum value is set slightly higher than the highest possibl...

Page 41: ...der Function pulse outputs Independent Occurrence of changes in direction of rotation If reverse pulses occur at the pulse inputs no pulses are output at the assigned pulse outputs Instead up to 10 re...

Page 42: ...e maximum output frequency e g pulse width 200 ms maximum frequency 2 5 Hz Note After the setting has been changed the electronic unit has to be restarted 3 13 Function relay 1 The relay function can...

Page 43: ...limit value is not exceeded the volumeter remains in bypass operation Therefore the relay is switched and thus the valve is closed after the Waiting period repeat bypass has expired If the limit valu...

Page 44: ...ly by repeated pressing of the key ARROW UP Menu 5 K factors Volumeter B The K factors of Volumeter B are entered in menu 5 K factors Volumeter B with rising frequency This maps the linearization char...

Page 45: ...the density table respectively 6 20 Density calculation For consumption measurement of heavy fuel oil engines the density calculation for fuel oils should always be used The density calculation is al...

Page 46: ...rs an alarm message is displayed This provides information used to eliminate the error The alarm is confirmed by pressing the SET key The alarm message disappears from the display Suitable measures fo...

Page 47: ...he frequencies in ascending order see Menu 4 K factors Volumeter A Page 44 The frequencies of Volumeter B are not entered in ascending order Enter the frequencies in ascending order see Menu 5 K facto...

Page 48: ...eliminate an electromagnetic interference using an oscilloscope e g terminate shield to Gnd At least one of the measured temperatures of the connected volumeter lies outside the permissible range Adap...

Page 49: ...error message is only displayed at the setting Function pulse inputs Encoder see 2 13 Function pulse inputs Page 38 Check the pulse inputs for channel A connections 7 18 a Square wave signal on both...

Page 50: ...ace the corresponding pick up Temperature sensor A or Temperature input A is defective Check the sensor connection Replace the sensor Check the temperature input Temperature sensor B or Temperature in...

Page 51: ...e of the pulse outputs Mode has been changed Correct the density table density calculation The maximum permissible flow rate was exceeded in Volumeter B Limit the flow rate Check the volumeter Use a l...

Page 52: ...correct analog output Correct the connection Negative flow rate The signal wires at the respective volumeter are connected incorrectly Swap the signal wires No flow or flow rate too low Check the alar...

Page 53: ...rrors are displayed 2 Eliminate the errors Overflow of the total value After an overflow of the total value the electronic unit displays the following for 3 decimal places 1 999 999 999 for 1 decimal...

Page 54: ...ersal mount fixing kit Fixing kit for pipe mounting mounting on OMG Fixing kit Application Article No suitable for Wall mounting UZA 20 BEM 300 BEM 500 Mounting M8 Fig 1 Mounting dimensions of univers...

Page 55: ...he control cabinet section 4 Slide on the sheeting 5 and sealing frame 6 and fasten using the screws and lock washers 7 Fixing kit Application Article No suitable for Mount on Volumeter OME UZA 21 1 2...

Page 56: ...Data Input voltage 100 240 V AC tolerance 93 265 V AC 47 63 Hz 135 370 V DC Starting current inrush 230 V AC 30 A max 115 V AC 15 A max limited by NTC in heated state higher Overvoltage protection at...

Page 57: ...ety Design to EN 60950 Degree of protection IP 20 Insulation voltage Input Output 3 kV routine tested EMC emitted interference EN 55011 B Immunity to interference EN 61000 6 2 Efficiency 83 depending...

Page 58: ...EEN 13 Input Data Power consumption 20 W Input voltage 90 264 V AC Frequency 47 63 Hz Current consumption 0 4 A max Leakage current 0 25 mA max Output Data Output voltage 24 V DC 2 Output current max...

Page 59: ...nalog input 2 30 mA max 316 mA Pulse input High 2 8 V max 30 V Low 0 4 V Max pulse frequency 10 kHz Input impedance 10 60 k Precision 0 1 at 4 20 mA 2 pulses Resolution 0 5 A 1 pulse Features BEA 80 B...

Page 60: ...us interface This remote display can display the electronic unit display values This means that the current consumption the total values and the temperatures can be displayed at distances of up to 200...

Page 61: ...sitiy calculation for diesel Temperature 70 C densitiy calculation for heavy fuel oil Density table Describes the relationship of density to temperature Thus maps the expansion coefficient Differentia...

Page 62: ...onological sequence Single line measurement The values of a volumeter are measured and evaluated Square wave signal Pulse signal with square wave form Supply line Line from the tank to the consumer Te...

Page 63: ...Operating Instructions OIE 12en Edition 2019 02 63 Notes Appendix Notes...

Page 64: ...www kral at KRAL AG 6890 Lustenau Austria Tel 43 5577 86644 0 E Mail kral kral at...

Reviews: