15
KPS-PA10 • Digital clamp meter
ENG
• Maximum voltage between terminals and earth ground: CAT II 1000V and
CAT III 600V
• Display: LCD, 1999 counts, updates 2-3 sec.
• Polarity indication: “―” displayed for negative polarity.
• Overrange indication: sólo la cifra “1” en la pantalla.
• Jaw opening capability: 42mm (max conductor size)
• Power: 9V battery, NEDA 1604 6F22 006P.
• Low battery indication: Symbol “
” appears on the display.
• Operating Environment: 0 to 40°C
• Storage temperature: -10 to +50°C
• Temperature coefficient: 0,1 x specified accuracy / °C (<18°C or > 28°C)
• Altitude: 2000 m
• Size: 250mm x 99mm x 43mm
• Weight: Approx. 460 g.
DC VOLTAGE
Range
Resolution
Accuracy
1000V
1V
±1,0% of reading ± 2 digits
Input Impedance: 10MΩ
AC VOLTAGE
Range
Resolution
Accuracy
700V
1V
±1,0% of reading ± 5 digits
Input Impedance: 10MΩ
Frequency range: 40Hz to 400Hz. Response: Average responding, calibrated
in rms. of a sine wave.
AC CURRENT
Range
Resolution
Accuracy
20A
0,01A
±2,0% of reading ± 5 digits
200A
0,1A
±2,0% of reading ± 5 digits
1000A
1A
±2,0% of reading ± 7 digits