40% of UT for 5 Periods
At 0º, 180º

Voltage

interruptions

IEC 61000-4-11

70% of UT for 25 Periods
At 0º, 180º

Table 5 – Signal input/output parts Port

Phenomenon

Basic EMC

standard

Immunity test levels

Professional healthcare facility

environment

Electrical fast transients/burst IEC 61000-4-4

±2 kV
5kHz repetition frequency

Conducted disturbances

induced by RF fields

IEC 61000-4-6

3V, 0.15MHz-80MHz
80%AM at 1kHz

Summary of Contents for KSOC Series

Page 1: ...OXYGEN CONCENTRATOR KSOC SERIES USER S MANUAL Read carefully before use Rev 1 4 ...

Page 2: ...decrease Knob Fragile handle with care Type BF applied part Keep dry Authorized representative in the European Community This side up Class II equipment ON power Batch code OFF power Date of Manufacturer Alternating current Manufacturer Operating instructions Marking of electrical and electronic equipment in accordance with 2012 19 EU WEEE ...

Page 3: ...in Material 8 6 Technical Parameters 9 7 Models and Performance 11 8 Working Conditions 11 9 Operation Instruction 12 10 Atomization for M N models 14 11 Timing and Remote Control 14 12 Alarm and Indicator System 15 13 Maintenance and Service 16 14 Disposable 18 15 Storage and Transportation 18 16 Troubleshooting 19 17 Service 20 18 EMC Guidelines 21 ...

Page 4: ...fore using the concentrator 2 If do not remove the bolts on the bottom of the oxygen concentrator the concentrator will not work well 3 Keep the bolts for next transportation The picture of the fixed bolts on the bottom of oxygen concentrator and tool as bellow The picture is only for reference please make the object as the standard ...

Page 5: ...he product near the heat resource or fire B The product is not appropriate to use in too humid environment such as bathroom During the operation ensure there are no humidification devices within 2 meters around and after cleaning filter components they must be totally dried before re use C Do not operate the product near flammable materials as grease oil detergent etc Neither uses such materials a...

Page 6: ... attention to electricity safety Do not turn on the product if the plug or power lines damaged and ensure to cut off the power when cleaning the concentrator or cleaning and replacing the filters C Install the regulator device when the voltage is higher than the normal range or in fluctuation D To extend the life span of the concentrator reboot 5 minutes after each shutdown to prevent the compress...

Page 7: ...sual use please substitute the filter sponge both in outlet and delivery with those provided or certified by our company J The cannula as accessory provided by our company is a sample for a try Use the cannula personally to keep it clean and health clean and disinfect it frequently it is suggested using disposal cannula each time K The temperature of application component including gas circuit is ...

Page 8: ...nd accessories we provided or we authenticated 2 Working Principle Flowchart 1 Working Principle KSOC series of Medical oxygen concentrator is made up of filter system compressor adsorb tower electrical control system humidifier system and the reasonable air course from case structure It adopts the current world s advanced transformation absorption PSA principle It separates the oxygen and nitroge...

Page 9: ...medical professional or trained patient Contraindications patients with oxygen poisoning and oxygen allergy are forbidden Oxygen concentrator mainly uses air as raw material and uses molecular sieve pressure swing adsorption process to produce oxygen with oxygen concentration ranging from 90 to 96 V V For medical institutions and other production of oxygen for hypoxia patients to use oxygen Warnin...

Page 10: ...ck View LCD Indicators and buttons Humidifier holder Power switch Humidifier Fixing elastic rope Atomization nozzle Handle Caster Humidifier connection tube Flow monitor Oxygen outlet Air filter Overload protector Power cable Caster Label ...

Page 11: ... Structure Material Filter system Foam filter AB Sresin Compressor ZL102cast alum filling PTFE Absorption chamber 6063 alum alloy O5 zeolite Electrical system PCB Silicone components Humidifying system ABS resin Polypropylene Case ABS resin ...

Page 12: ...Weight Kg Atomization Amount mL min Timing period Remote control distance KSOC 3 3 93 3 53 380X320X590 320 19 N M 10h N M KSOC 3N 0 2 10h N M KSOC 3T N M 10h 50m KSOC 3M 0 2 10h 50m KSOC 5 5 54 480 18 3 N M 10h N M KSOC 5N 0 2 10h N M KSOC 5T N M 10h 50m KSOC 5M 0 2 10h 50m KSOC 8 8 N M 10h N M KSOC 8N 0 2 10h N M KSOC 8T N M 10h 50m KSOC 8M 0 2 10h 50m ...

Page 13: ...m 2000m 3000m 3000m 4000m KSOC 3 90 80 75 65 55 KSOC 5 90 80 75 63 53 KSOC 8 90 80 70 60 50 Outlet pressure has effect to oxygen density When the outlet pressure is Zero the standard flow is KSOC 3 series is 3L min KSOC 5 series is 5L min KSOC 8 series is 8L min When the outlet pressure is 7kpa the standard flow is KSOC 3 series is 2 7L min KSOC 5 series is 4 5L min KSOC 8 series is 7 2L min Funct...

Page 14: ...d power off alarm Class of safety type It belongs to CLASS II equipment Type BF Applied part for safety type Oxygen concentrator is non AP APG device Protection class of case IPX0 Running mode continuous operation Outlet pressure 0 04MPa 0 07MPa Pressure range of atomization outlet 60kPa 250kPa Safety valve pressure range 15kPa 40kPa Safety valve pressure range 250kPa 280kPa Overloading protector ...

Page 15: ...oncentrator power cord to the socket of the power supply AC 220 240V 50 60HZ or AC110V 60HZ and connect the power 3 Unscrew the humidifier pour the distilled water or cold water into the humidifier to its prescribed level from Max and Min level in case of over flow or uneven humidifier for lack of water as below figure shows Maximum water level Water level Minimum water level Power off alarm indic...

Page 16: ...timing mode the symbol of after timing shutdown displays that the automatically shutdown has not been started the figures after the timing shutdown show the remaining time before timing shutdown Press the keys of and on the panel and the time of timing shutdown could be increased or reduced The indicator code will show in the case of occurrence of abnormal situation of oxygen concentrator for the ...

Page 17: ...he oxygen concentrator 0 2mL min 11 Timing and Remote Control 1 Timing operation KSOC oxygen concentrator can be shut down with timing equipment The longest period is 10 hours The timing interval can be 10 minutes within 1 hour timing or 30 minutes more than 1 hour timing When the shut hour is set the system comes into count down time and the oxygen concentrator LCD will shows the remaining time W...

Page 18: ...ows E05 the air pressure of the inside tube 260kPa or the magnetic valve anomaly c The indicator lights yellow and the fault code shows E02 the air pressure of the inside tube 20kPa or the pipe shedding Please turn off the concentrator immediately and check whether the in outlet of the oxygen concentrator is blocked or not Keep the in outlet clean and then restart the concentrator If there still h...

Page 19: ...t fleece to wipe the case of the oxygen concentrator Please cut off the power to ensure safety 2 Cleaning of the humidifier Please empty the humidifier each time after using the equipment Wash the humidifier every day a Pull out the connection soft tube and take out the humidifier b Turn the lid open Wash it with water You can use a clean brush or towel to clean it if there is any besmirch You can...

Page 20: ...ve the used inner intake air filter to replace a new one as shown in below figures 4 The replacement of over load protector When suspect or identify over load protector works power Link normal Start up power outage alarms press the reset button in the scuttle of the oxygen concentrator back shield The overload protector locates on the back of concentrator as shown in below figure 5 Replacement of ...

Page 21: ... affirmative and negative pole 14 Disposable Please dispose of the disposable cannula filter sponge and defective oxygen concentrator according to local relative laws in case of environmental pollution 15 Storage and Transportation 1 The unit should be stored 40 55 Relative humidity 93 Pressure 50kPa 106kPa in the nice environment without the corrosive air and with a good wind 2 Be careful to tran...

Page 22: ...he requested current capacity 1 Nose sucker jammed or damaged 2 Mask jammed or damaged 3 Wet bottles jammed or damaged 4 Oxygen tubes surpass the stipulation or have the bend 1 Change or rectify the oxygen tubes 2 Take down the wet bottle clean or change it If it s not in the above cases and there is still no oxygen output please contact the distributor or the manufacturer Screen prompt code and p...

Page 23: ...y provides the parts to repair with reasonable charge if it could not be used The following conditions are not included within the warranty coverage 1 damage and deformation caused by crush 2 damage caused by water and rain 3 abnormal working state caused by self dismantle of user Notes 1 Please take care of the purchase invoice and maintenance card for service 2 The non controllable factor or the...

Page 24: ...ility environment Essential Performance Noise KSOC 3 Series 53dB KSOC 5 KSOC 8 Series 54dB O2 density V V 93 3 WARNING Portable RF communications equipment including peripherals such as antenna cables and external antennas should be used no closer than 30 cm 12 inches to any part of the Oxygen Concentrator including cables specified by the manufacturer Otherwise degradation of the performance of t...

Page 25: ...itters and cables other than those specified by Konsung with the exception of accessories and cables sold by Konsung of Oxygen Concentrator as replacement parts for internal components may result in increased Emissions or decreased Immunity of the Oxygen Concentrator WARNING Use of this equipment Oxygen Concentrator adjacent to or stacked with other equipment should be avoided because it could res...

Page 26: ...relocating or re orienting the equipment Table 2 Enclosure Port Phenomenon Basic EMC standard Immunity test levels Professional healthcare facility environment Electrostatic Discharge IEC 61000 4 2 2kV 4kV 6kV contact 2kV 4kV 8kV air Radiated RF EM field IEC 61000 4 3 3V m 80MHz 2 5GHz 80 AM at 1kHz Proximity fields from RF wireless communications equipment IEC 61000 4 3 Refer to table 3 Rated pow...

Page 27: ... Table 4 Input AC power Port Phenomenon Basic EMC standard Immunity test levels Professional healthcare facility environment Electrical fast transients burst IEC 61000 4 4 2 kV 5kHz repetition frequency Surges Line to line IEC 61000 4 5 0 5 kV 1 kV Surges Line to ground IEC 61000 4 5 Do not apply Conducted disturbances induced by RF fields IEC 61000 4 6 3V 0 15MHz 80MHz 80 AM at 1kHz Voltage dips ...

Page 28: ...180º Table 5 Signal input output parts Port Phenomenon Basic EMC standard Immunity test levels Professional healthcare facility environment Electrical fast transients burst IEC 61000 4 4 2 kV 5kHz repetition frequency Conducted disturbances induced by RF fields IEC 61000 4 6 3V 0 15MHz 80MHz 80 AM at 1kHz ...

Reviews: