23
Harmonic distortion
IEC 61000-3-2
Class A
100Hz to 2kHz,
2.5minutes
Voltage fluctuations
and flicker
IEC 61000-3-3
2 hours, Clause 5
NOTE: The Emissions characteristics of this equipment make it suitable
for use in industrial areas and hospitals (CISPR 11 class A). If it is used in
a residential environment (for which CISPR 11 class B is normally
required) this equipment might not offer adequate protection to
radio-frequency communication services. The user might need to take
mitigation measures, such as relocating or re-orienting the equipment.
Table 2 - Enclosure Port
Phenomenon
Basic EMC
standard
Immunity test levels
Professional healthcare facility
environment
Electrostatic
Discharge
IEC 61000-4-2 ±2kV, ±4kV, ±6kV contact
±2kV, ±4kV, ±8kV air
Radiated RF EM field
IEC 61000-4-3
3V/m
80MHz-2.5GHz
80% AM at 1kHz
Proximity fields from RF
wireless communications
equipment
IEC 61000-4-3
Refer to table 3
Rated power frequency
magnetic fields
IEC 61000-4-8
3A/m
50Hz
&
60Hz
Table 3 – Proximity fields from RF wireless communications equipment
Test frequency
(MHz)
Band
(MHz)
Immunity test levels
Professional healthcare facility environment
385
380-390
Pulse modulation 18Hz, 27V/m
450
430-470
FM, ±5kHz deviation, 1kHz sine, 28V/m