[ViZion DR+ 1417V2 Wireless Panel Calibration Manual]
[Ultra 4.2.9.0]
16
Customer Support
1.800.366.5343
–
11. Continue to take exposures with
techniques set to
70kV / 7mAs
for the
remaining frames (5-19).
12. After taking all 19 defect exposures,
click
Back
, if you want to retake any
previous exposures, then click
Prep
and
retake the exposure. Otherwise, if you
are satisfied with all exposures, click
Next
in the top right corner.
13. Wait for the
Defect Map Generation
to
complete and then click
Next
.
14. Calibration is complete.
Close
the
window by clicking
X
in top right corner.